Skip to main content
. 2010 Sep 1;11(4):130–143. doi: 10.1120/jacmp.v11i4.3164

Table 1.

MapCHECK – Eclipse comparison.

CW Coronal CW Sagittal CCW Coronal CCW Sagittal
Average % agreem. Average % agreem. Average % agreem. Average % agreem.
DTA
Gamma 93.5%
95.4% 96.5%
97.9% 94.0%
95.7% 93.7%
96.2%
CAX Average % discrepancy CAX Average % discrepancy CAX Average % discrepancy CAX Average % discrepancy
CAX [meas.‐calc.]/calc*100(%) ‐3.5% ‐3.5% ‐2.2% ‐2.8%

Percentage of diodes receiving a dose greater than 10% of the maximum dose that met the acceptance criteria (3%/3 mm) for DTA and gamma index approaches, and percent discrepancy of the CAX diode.