Table 1.
CW Coronal | CW Sagittal | CCW Coronal | CCW Sagittal | |
---|---|---|---|---|
Average % agreem. | Average % agreem. | Average % agreem. | Average % agreem. | |
DTA Gamma | 93.5% 95.4% | 96.5% 97.9% | 94.0% 95.7% | 93.7% 96.2% |
CAX Average % discrepancy | CAX Average % discrepancy | CAX Average % discrepancy | CAX Average % discrepancy | |
CAX | ‐3.5% | ‐3.5% | ‐2.2% | ‐2.8% |
Percentage of diodes receiving a dose greater than 10% of the maximum dose that met the acceptance criteria (3%/3 mm) for DTA and gamma index approaches, and percent discrepancy of the CAX diode.