Table 1.
Same Floor | Distance to Non‐controlled Occupied Area (cm) | ||||
---|---|---|---|---|---|
Lead (mm) | 200 | 300 | 400 | 500 | 600 |
0 | 20 | 40 | 80 | 120 | 170 |
3 | 30 | 60 | 110 | 170 | 240 |
6 | 40 | 90 | 150 | 240 | 340 |
10 | 50 | 100 | 180 | 280 | 410 |
Same Floor | Distance to Non‐controlled Occupied Area (cm) | ||||
---|---|---|---|---|---|
Lead (mm) | 200 | 300 | 400 | 500 | 600 |
0 | 20 | 40 | 80 | 120 | 170 |
3 | 30 | 60 | 110 | 170 | 240 |
6 | 40 | 90 | 150 | 240 | 340 |
10 | 50 | 100 | 180 | 280 | 410 |