Skip to main content
. 2010 Sep 7;11(4):263–273. doi: 10.1120/jacmp.v11i4.3151

Table 2.

Workload limits in Gy/week for the occupied areas in the floor below the Mobetron as a function of concrete thickness. All calculations were made assuming an occupancy factor of unity.

Floor Below Distance to Non‐controlled Occupied Area (cm)
Concrete (cm) 200 300 400 500 600
15 30 70 130 200 290
25 50 120 220 340 490