Skip to main content
. 2010 Aug 5;11(4):222–237. doi: 10.1120/jacmp.v11i4.3172

Table 1.

The individual parameters for different fitting methods and different irradiation geometries.

Geometry 1 2 3
SSD (cm) 90 95 90
Field size (cm2)
20×20
10×10
5×5
Depth (cm) 10 5 10
Fitting parameters with one background and eight dose points
P0
247 240 254
Ps
3585 3577 3209
m 30.80 33.58 40.18
Method I: Universal Ps and P0
Universal P0 245 245 245
Universal Ps 3542 3542 3542
Ia (one point a ): m 31.28 33.37 37.62
Ib (two points b ): m 31.17 33.48 38.28
Method II: Universal
Ps
P0
247 240 254
Universal Ps 3542 3542 3542
IIa (one point a ): m 31.23 33.50 37.36
IIb (two points b ): m 31.11 33.63 37.98
Method III
P0
247 240 254
IIIa (two points c ): Ps 3558 3494 3200
IIIb (three points d ): Ps 3538 3506 3200
IIIa (two points c ): m 30.91 33.95 40.19
IIIb (three points d ): m 31.10 33.80 40.16
a

Dose point: ~80 cGy

b

Dose points: ~40 and ~80 cGy

c

Dose points: ~40 and ~120 cGy