Passing rates between MapCHECK and TPS calculated dose distributions and MatriXX and TPS calculated dose distributions for three 6 MV IMRT fields and three 18 MV IMRT fields. Both the original TPS calculated dose distributions and the TPS calculated dose distributions convolved with a Gaussian with a σ of 2.5 mm were used to compare with the MatriXX measured dose distributions.

MapCHECK with TPS MatriXX with TPS (Original) MatriXX with TPS (Convolved)
3%/3mm
2%/2mm
3%/3mm
2%/2mm
3%/3mm
2%/2mm
6 MV Field A 98.3% 94.4% 98.1% 93.3% 100.0% 96.1%
6 MV Field B 100.0% 96.8% 97.6% 90.5% 99.2% 94.8%
6 MV Field C 99.6% 96.1% 97.9% 92.7% 98.8% 95.3%
Average (6 MV) 99.3% 95.8% 97.9% 92.2% 99.3% 95.4%
18 MV Field A 97.8% 94.9% 96.3% 91.4% 96.9% 94.3%
18 MV Field B 95.3% 93.2% 93.6% 89.8% 95.2% 93.9%
18 MV Field C 99.7% 96.1% 95.8% 90.7% 97.6% 95.1%
Average (18 MV) 97.6% 94.7% 95.2% 90.6% 96.6% 94.4%