Skip to main content
. 2009 Feb 5;10(1):120–128. doi: 10.1120/jacmp.v10i1.2915

Table 1.

P‐values from two‐tailed Wilcoxon rank‐sum test. The test compared the distribution of passing rates before and after the introduction of Mlc position errors.

Random 1 mm p‐Value
1mm
2 mm ‐2 mm
EBT Film DTA
2%/2mm
3.28E‐01 9.59E‐01 1.05E‐01 2.81E‐02 6.22E‐04
3%/3mm
5.56E‐01 7.98E‐01 1.61E‐01 7.74E‐02 1.10E‐03
Gamma
2%/2mm
5.05E‐01 9.81E‐01 1.30E‐01 2.81E‐02 6.22E‐04
3%/3mm
6.29E‐01 1.00E+00 3.14E‐01 9.95E‐02 1.90E‐03
MapCHECK DTA
2%/2mm
1.10E‐02 1.29E‐05 2.60E‐03 2.35E‐11 6.86E‐10
3%/3mm
3.20E‐02 1.90E‐05 3.60E‐03 2.84E‐11 1.62E‐08
Gamma
2%/2mm
1.76E‐02 1.08E‐05 3.45E‐04 3.97E‐11 3.28E‐10
3%/3mm
7.59E‐05 3.85E‐06 3.57E‐04 4.33E‐12 9.20E‐10