Skip to main content
. 2009 Sep 30;10(4):241–251. doi: 10.1120/jacmp.v10i4.2923

Table 3.

Evaluation of ionization chambers for DMLC QA measurements.

Average (range) (cGy)
Average±2SD(cGy)
Chamber FSPB Algorithm Measurement
(Dw,FSPB/Dw,meas)DMLC
PinPoint 200.9 (101.7, 277.1) 211.5 (109.8, 302.5)
0.944±0.070
IC10 200.5 (103.1, 277.1) 208.5 (108.7, 293.1)
0.963±0.048
Farmer‐type 200.1 (105.0, 276.9) 207.3 (110.1, 287.2)
0.973±0.044

The average dose calculation and measurement are presented in columns two and three. The average dose ratios ±2SD are presented in column four.