. 2009 Sep 30;10(4):64–78. doi: 10.1120/jacmp.v10i4.3080

Table 1.

Step‐and‐shoot IMRT: ion chamber dose error and y analysis for all detectors receiving 30% of the “normalization” dose.

Plan Plan Type 6 MV 15 MV
IC Dose‐Difference, % γ(3%, 3mm) IC Dose‐Difference, % γ(3%, 3mm)
1 Phantom (Prostate)
0.2
100
2 Prostate 0.9 99.8 1.2 100
3 H&N
0.9
100 1.2 99.8
4 H&N
0.8
100
0.2
100
5 H&N (split beam) 0.2 92.0 0.8 95.4
6 H&N (split beam)
0.6
99.6 0.8
7 Abdomen (split beam) 0.5 95 1.5