Table 4.

Comparison of wedge factors between water and Plastic Water a Diagnostic Therapy (DT) for a 6‐MV beam

Field size (cm) Plastic Water DT Water
15‐Degree wedge
10×10
0.509 0.499
20×20
0.521 0.522
45‐Degree wedge
10×10
0.711 0.714
20×20
0.724 0.723
a

CIRS, Norfolk, VA.