. 2008 Apr 29;9(2):83–97. doi: 10.1120/jacmp.v9i2.2702

Figure 6.

Figure 6

(a) Subtraction of BEAM (BEAMnrc code, National Research Council of Canada implementation) dose from FMC‐NX [fast Monte Carlo code, NumeriX LLC (New York, NY) implementation] dose. Dose calculated using an 18‐MV beam set to a 10×10 cm field. Displayed dose ranging from 8.7% (0.07% cGy per monitor unit) to +8.7% (+0.07% cGy per monitor unit). The red line indicates where the profile for Fig. 8 was taken. (b) Gamma values taken from the boxed region shown in (a), with 3% and 3 mm parameters and 7640 of 7725 voxels having a γ of 1 or less.