. 2008 Jul 9;9(4):17–36. doi: 10.1120/jacmp.v9i4.2781

Table 2.

Comparison of the CT phantom translational shifts and CT/CT registration shifts. The phantom “skin” was used as the 3D volumetric image registration landmark, and the visual color homogeneity was used as registration criterion. The registration shifts were calculated by physical distance (mm)=(voxel shift)×(voxel size), where 1.0voxel=0.78mm. The registration was cross‐confirmed using the “skull” as landmark. The uncertainty of the CT couch positioning was specified within ±0.5mm in longitudinal direction and within ±0.1mm in lateral direction by the manufacturer. The phantom positioning had ±0.1mm uncertainty by aligning reference lines of 0.13 mm in width under optical amplification.

Experimental Shifts (mm) Registration Shifts (mm) Statistical Analysis (mm)
Lateral
XExp
X1
X2
X3
X4
XAvg
XExpXAvg St. Dev.
5.0±0.1
4.92 4.92 4.99 5.07 4.98 0.02 0.08
10.0±0.1
9.92 10.14 9.99 9.99 10.01
0.01
0.09
15.0±0.1 Average 14.91 14.91 14.91 15.08 14.95 0.05 0.10
Average 0.02 0.09
Longitudinal
ZExp
Z1
Z2
Z3
Z4
ZAvg
ZExpZAvg
St. Dev.
5.0±0.5
5.00 4.68 4.61 4.68 4.74 0.26 0.04
10.0±0.5
9.76 9.53 9.92 9.59 9.70 0.30 0.21
15.0±0.5
14.84 14.52 15.15 14.44 14.74 0.26 0.39
Average 0.27 0.21