Table 1.

Small plastic target isocentric positioning test a

Displacement direction Mean Standard deviation Min Max 95% CI for mean
LR (ΔX) 0.19 0.16
0.05
0.50 0.06
SI (ΔY) 0.48 0.22 0.09 0.82 0.09
AP (ΔZ) 0.76 0.17 0.40 1.00 0.07
3D Vector 0.95 0.18 0.67 1.28 0.07
a

N=27; all values in millimeters.

CI=confidence interval; LR=leftright; SI=superiorinferior; AP=anteriorposterior; 3D=threedimensional.