. 2006 Nov 28;7(4):22–32. doi: 10.1120/jacmp.v7i4.2278

Table 4.

Uncertainty of the MOSFET dose for IMRT therapy given as 1 SD (high sensitivity)

Uncertainty (%)
Step 1: MOSFET calibration (Ref Cond)
Dose measured with ionization chamber 1.0%
Reading of the MOSFET detector 1.3%
Reproducibility of the setting 0.5%
Combined uncertainty of step 1 1.7%
Step 2: Correction factors
SSD correction factor 1.4%
CR value 1.5%
IMRT correction factor–MOSFET reading on solid water phantom 1.3%
Combined uncertainty of step 2 2.4%
Step 3: Measurement on the patient
Reading of the MOSFET detector 1.3%
Positioning (2% over 2 mm radius) 1.5%
MOSFET detector long‐term stability over time 3.0%
Combined uncertainty of step 3 3.6%
Combined uncertainty of steps
1+2+3
4.6%