Quality control tests for medical linear accelerators (tolerances and action levels are specified in millimeters unless otherwise stated)

Designator Test Performance
Tolerance Action
Daily
DL1 Door interlock/last person out Functional
DL2 Motion interlock Functional
DL3 Couch brakes Functional
DL4 Beam status indicators Functional
DL5 Patient audiovisual monitors Functional
DL6 Room radiation monitors Functional
DL7 Beam interrupt/counters Functional
DL8 Lasers/crosswires 1 2
DL9 Optical distance indicator 1 2
DL10 Optical back pointer 2 3
DL11 Field size indicator 1 2
DL12 Output constancy—photons 2% 3%
DL13 Dynamic wedge factors 1% 2%
DL14 Output constancy—electrons 2% 3%
Monthly
ML1 Emergency off Functional
ML2 Wedge, tray cone interlocks Functional
ML3 Accessories integrity and centering Functional
ML4 Gantry angle readouts 0.5°
ML5 Collimator angle readouts 0.5°
ML6 Couch position readouts 1 2
ML7 Couch isocenter 1 2
ML8 Couch angle 0.5°
ML9 Optical distance indicator 1 2
ML10 Crosswire centering 1 2
ML11 Light/radiation coincidence 1 2
ML12 Field size indicator 1 2
ML13 Relative dosimetry 1% 2%
ML14 Central axis depth dose reproducibility 1 (2%) 2 (3%)
ML15 Beam flatness 2% 3%
ML16 Beam symmetry 2% 3%
ML17 Records Complete
Annually
AL1 Reference dosimetry—TG51 1% 2%
AL2 Relative output factor reproducibility 1% 2%
AL3 Wedge transmission factor reproducibility 1% 2%
AL4 Accessory transmission factor reproducibility 1% 2%
AL5 Output reproducibility vs. gantry angle 1% 2%
AL6 Beam symmetry reproducibility vs. gantry angle 2% 3%
AL7 Monitor chamber linearity 1% 2%
AL8 End monitor effect 0.1 MU 0.2 MU
AL9 Collimator rotation isocenter 1 2
AL10 Gantry rotation isocenter 1 2
AL11 Couch rotation isocenter 1 2
AL12 Coincidence of collimator, gantry, couch axes 1 2
AL13 Coincidence of isocenters 1 2
AL14 Couch deflection 3 5
AL15 Independent quality control review Complete