Table 2.
Quality control tests for conventional simulators (tolerances and action levels are specified in millimeters unless otherwise stated)
Designator | Test | Performance | ||
---|---|---|---|---|
Tolerance | Action | |||
Daily | ||||
DS1 | Door interlock | Functional | ||
DS2 | Motion interlock | Functional | ||
DS3 | Beam status indicators | Functional | ||
DS4 | Emergency off buttons | Functional | ||
DS5 | Collision avoidance | Functional | ||
DS6 | Lasers/crosswires | 1 | 2 | |
DS7 | Optical distance indicator | 1 | 2 | |
DS8 | Crosswires/reticle/block tray | 1 | 2 | |
DS9 | Light/radiation coincidence | 1 | 2 | |
DS10 | Field size indicators | 1 | 2 | |
Monthly | ||||
MS1 | Gantry angle readouts | 0.5° | 1° | |
MS2 | Collimator angle readouts | 0.5° | 1° | |
MS3 | Couch position readouts | 1 | 2 | |
MS4 | Alignment of FAD movement | 1 | 2 | |
MS5 | Couch isocenter | 2 | 3 | |
MS6 | Couch parallelism | 1 | 2 | |
MS7 | Laser/crosswire isocentricity | 1 | 2 | |
MS8 | Optical distance indicator | 1 | 2 | |
MS9 | Crosswire centering | 1 | 2 | |
MS10 | Light/radiation coincidence | 1 | 2 | |
MS11 | Field size indicators | 1 | 2 | |
MS12 | Records | Complete | ||
Six‐monthly | ||||
SS1 | Lead apron | Functional | ||
SS2 |
|
5% | 10% | |
SS3 | Reference dosimetry | 5% | 10% | |
SS4 | Beam quality (HVL) | 5% | 10% | |
SS5 | Automatic exposure control | 5% | 10% | |
SS6 | Focal spot | Reproducible | ||
SS7 | Contrast | Reproducible | ||
SS8 | Resolution | Reproducible | ||
SS9 | Fluoroscopic timer | 5% | 10% | |
Annually | ||||
AS1 | Redefine isocenter | 1 | 2 | |
AS2 | Couch deflection | 3 | 5 | |
AS3 | Alignment of focal spots | 0.5 | 1 | |
AS4 | Independent quality control review | Complete |