Daily
|
|
|
|
DS1 |
Door interlock |
Functional |
DS2 |
Motion interlock |
Functional |
DS3 |
Beam status indicators |
Functional |
DS4 |
Emergency off buttons |
Functional |
DS5 |
Collision avoidance |
Functional |
DS6 |
Lasers/crosswires |
1 |
2 |
DS7 |
Optical distance indicator |
1 |
2 |
DS8 |
Crosswires/reticle/block tray |
1 |
2 |
DS9 |
Light/radiation coincidence |
1 |
2 |
DS10 |
Field size indicators |
1 |
2 |
Monthly
|
|
|
|
MS1 |
Gantry angle readouts |
0.5° |
1° |
MS2 |
Collimator angle readouts |
0.5° |
1° |
MS3 |
Couch position readouts |
1 |
2 |
MS4 |
Alignment of FAD movement |
1 |
2 |
MS5 |
Couch isocenter |
2 |
3 |
MS6 |
Couch parallelism |
1 |
2 |
MS7 |
Laser/crosswire isocentricity |
1 |
2 |
MS8 |
Optical distance indicator |
1 |
2 |
MS9 |
Crosswire centering |
1 |
2 |
MS10 |
Light/radiation coincidence |
1 |
2 |
MS11 |
Field size indicators |
1 |
2 |
MS12 |
Records |
Complete |
Six‐monthly
|
|
|
|
SS1 |
Lead apron |
Functional |
SS2 |
|
5% |
10% |
SS3 |
Reference dosimetry |
5% |
10% |
SS4 |
Beam quality (HVL) |
5% |
10% |
SS5 |
Automatic exposure control |
5% |
10% |
SS6 |
Focal spot |
Reproducible |
SS7 |
Contrast |
Reproducible |
SS8 |
Resolution |
Reproducible |
SS9 |
Fluoroscopic timer |
5% |
10% |
Annually
|
|
|
|
AS1 |
Redefine isocenter |
1 |
2 |
AS2 |
Couch deflection |
3 |
5 |
AS3 |
Alignment of focal spots |
0.5 |
1 |
AS4 |
Independent quality control review |
Complete |