Daily
|
|
|
DK1 |
Patient monitoring audiovisual devices |
Functional |
DK2 |
Door closing mechanism and interlock |
Functional |
DK3 |
Couch movement and brakes |
Functional |
DK4 |
Unit motions and motion stops |
Functional |
DK5 |
Interlocks for added filters/kV‐filter choice |
Functional |
DK6 |
Beam status indicators |
Functional |
DK7 |
Beam‐off at key‐off test |
Functional |
DK8 |
Emergency off test |
Functional |
DK9 |
kV and mA indicators |
Functional |
DK10 |
Backup timer/monitor unit channel check |
1% |
2% |
DK11 |
Dosimetric test: output check |
3% |
5% |
Monthly
|
|
|
MK1 |
Mechanical stability and safety |
Functional |
MK2 |
Cone selection and competency |
Functional |
MK3 |
Physical distance indicators |
2 |
3 |
MK4 |
Accuracy of head tilt and rotation readouts |
1° |
1.5° |
MK5 |
Light/x‐ray field coincidence |
2 |
3 |
MK6 |
Light field size |
2 |
3 |
MK7 |
X‐ray field size indicator |
2 |
3 |
MK8 |
X‐ray field uniformity/filter integrity |
5% |
8% |
MK9 |
Timer and end effect error |
Characterize |
min |
MK10 |
Output linearity |
1% |
MK11 |
Output reproducibility |
Characterize |
<.03 CoV |
MK12 |
Beam quality |
10% |
15% |
MK13 |
Output calibration verification |
2% |
3% |
MK14 |
Timer accuracy verification |
2% |
3% |
MK15 |
Records |
Complete |
Annually
|
|
|
AK1 |
Reference dosimetry |
1% |
2% |
AK2 |
Alignment of focal spots |
0.5 |
1 |
AK3 |
kVp measurement |
5% |
10% |
AK4 |
Focal spot size |
Reproducible |
AK5 |
Independent quality control review |
Complete |