Quality control tests for C60o teletherapy units (tolerances and action levels are specified in millimeters unless otherwise stated)

Designator Test Performance
Tolerance Action
Daily
DCO1 Door interlock/last person out Functional
DCO2 Motion interlock Functional
DCO3 Couch brakes Functional
DCO4 Beam status indicators Functional
DCO5 Patient audiovisual monitors Functional
DCO6 Room radiation monitors Functional
DCO7 Emergency off Functional
DCO8 Beam interrupt/counters Functional
DCO9 Head swivel lock Functional
DCO10 Lasers/crosswires 1 2
DCO11 Optical distance indicator 1 2
DCO12 Optical back pointer 2 3
DCO13 Field size indicator 1 2
Monthly
MCO1 Latching of wedges, trays Functional
MCO2 Wedge interlocks Functional
MCO3 Gantry angle readouts 0.5°
MCO4 Collimator angle readouts 0.5°
MCO5 Couch position readouts 1 2
MCO6 Couch rotation isocenter 2 3
MCO7 Optical distance indicator 1 2
MCO8 Crosswire centering 1 2
MCO9 Light/Radiation coincidence 2 3
MCO10 Field size indicator 1 2
MCO11 Relative dosimetry 1% 2%
MCO12 Shutter error Reproducible
MCO13 Beam symmetry (source position) 2% 3%
MCO14 Records Complete
Annually
ACO1 Reference dosimetry 1% 2%
ACO2 Relative output factor reproducibility 1% 2%
ACO3 Central axis depth dose reproducibility 1% 2%
ACO4 Wedge transmission factor reproducibility 1% 2%
ACO5 Accessory transmission factor reproducibility 1% 2%
ACO6 Output reproducibility vs. gantry angle 1% 2%
ACO7 Beam symmetry reproducibility vs. gantry angle 2% 3%
ACO8 Timer linearity 1% 2%
ACO9 Shutter error 0.03 min. 0.05 min.
ACO10 Collimator rotation isocenter 2 3
ACO11 Gantry rotation isocenter 2 3
ACO12 Couch rotation isocenter 2 3
ACO13 Coincidence of collimator, gantry, couch axes 2 3
ACO14 Coincidence of isocenters 2 3
ACO15 Couch deflection 3 5
ACO16 Independent quality control review Complete