Daily
|
|
|
DCO1 |
Door interlock/last person out |
Functional |
DCO2 |
Motion interlock |
Functional |
DCO3 |
Couch brakes |
Functional |
DCO4 |
Beam status indicators |
Functional |
DCO5 |
Patient audiovisual monitors |
Functional |
DCO6 |
Room radiation monitors |
Functional |
DCO7 |
Emergency off |
Functional |
DCO8 |
Beam interrupt/counters |
Functional |
DCO9 |
Head swivel lock |
Functional |
DCO10 |
Lasers/crosswires |
1 |
2 |
DCO11 |
Optical distance indicator |
1 |
2 |
DCO12 |
Optical back pointer |
2 |
3 |
DCO13 |
Field size indicator |
1 |
2 |
Monthly
|
|
|
MCO1 |
Latching of wedges, trays |
Functional |
MCO2 |
Wedge interlocks |
Functional |
MCO3 |
Gantry angle readouts |
0.5° |
1° |
MCO4 |
Collimator angle readouts |
0.5° |
1° |
MCO5 |
Couch position readouts |
1 |
2 |
MCO6 |
Couch rotation isocenter |
2 |
3 |
MCO7 |
Optical distance indicator |
1 |
2 |
MCO8 |
Crosswire centering |
1 |
2 |
MCO9 |
Light/Radiation coincidence |
2 |
3 |
MCO10 |
Field size indicator |
1 |
2 |
MCO11 |
Relative dosimetry |
1% |
2% |
MCO12 |
Shutter error |
Reproducible |
MCO13 |
Beam symmetry (source position) |
2% |
3% |
MCO14 |
Records |
Complete |
Annually
|
|
|
ACO1 |
Reference dosimetry |
1% |
2% |
ACO2 |
Relative output factor reproducibility |
1% |
2% |
ACO3 |
Central axis depth dose reproducibility |
1% |
2% |
ACO4 |
Wedge transmission factor reproducibility |
1% |
2% |
ACO5 |
Accessory transmission factor reproducibility |
1% |
2% |
ACO6 |
Output reproducibility vs. gantry angle |
1% |
2% |
ACO7 |
Beam symmetry reproducibility vs. gantry angle |
2% |
3% |
ACO8 |
Timer linearity |
1% |
2% |
ACO9 |
Shutter error |
0.03 min. |
0.05 min. |
ACO10 |
Collimator rotation isocenter |
2 |
3 |
ACO11 |
Gantry rotation isocenter |
2 |
3 |
ACO12 |
Couch rotation isocenter |
2 |
3 |
ACO13 |
Coincidence of collimator, gantry, couch axes |
2 |
3 |
ACO14 |
Coincidence of isocenters |
2 |
3 |
ACO15 |
Couch deflection |
3 |
5 |
ACO16 |
Independent quality control review |
Complete |