Table 5.

Quality control tests for multileaf collimators (tolerances and action levels are specified in millimeters unless otherwise stated)

Designator Test Performance
Tolerance Action
Patient‐specific
PM1 Verification of transferred data vs. printed template 1 2
PM2 Daily verification of correct data Reproducibility
PM3 Verification of record and verify programming Reproducibility
Monthly
MM1 Digitizer check (if used) Functional
MM2 Light and radiation field coincidence 1 2
MM3 Leaf positions for standard field template 1 2
MM4 Electron field interlocks Functional
MM5 a Leaf alignment 1
MM6 Records Complete
Yearly
AM1 Leaf transmission (all energies) Reproducibility
AM2 Leakage between leaves (all energies) Reproducibility
AM3 a Transmission through abutting leaves Reproducibility
AM4 Stability with gantry rotation Reproducibility
AM5 Alignment with jaws 1
AM6 Independent quality control review Complete
a

May not apply to all designs.