Table 6.
Quality control tests for electronic portal imaging devices (tolerances and action levels are specified in millimeters unless otherwise stated)
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Daily | |||
| DE1 | Mechanical integrity | Functional | |
| DE2 | Electrical integrity | Functional | |
| DE3 | Collision interlocks | Functional | |
| DE4 | Image quality | Reproducibility | |
| Monthly | |||
| ME1 | Positioning in the imaging plane | 1 | 2 |
| ME2 | Positioning perpendicular to the imaging plane | 10 | 20 |
| ME3 | Image quality | Reproducibility | |
| ME4 | Artifacts | Reproducibility | |
| ME5 | Spatial distortion | 1 | 2 |
| ME6 | Monitor controls | Reproducibility | |
| ME7 | Records | Complete | |
| Six monthly | |||
| SE1 | Spatial resolution | Reproducibility | |
| SE2 | Noise | Reproducibility | |
| SE3 | On‐screen measurement tools | 0.5 | 1 |
| SE4 | Setup verification tools | 0.5 (0.5°) | 1 (1°) |
| Annually | |||
| AE1 | Independent quality control review | Complete | |