Table 6.

Quality control tests for electronic portal imaging devices (tolerances and action levels are specified in millimeters unless otherwise stated)

Designator Test Performance
Tolerance Action
Daily
DE1 Mechanical integrity Functional
DE2 Electrical integrity Functional
DE3 Collision interlocks Functional
DE4 Image quality Reproducibility
Monthly
ME1 Positioning in the imaging plane 1 2
ME2 Positioning perpendicular to the imaging plane 10 20
ME3 Image quality Reproducibility
ME4 Artifacts Reproducibility
ME5 Spatial distortion 1 2
ME6 Monitor controls Reproducibility
ME7 Records Complete
Six monthly
SE1 Spatial resolution Reproducibility
SE2 Noise Reproducibility
SE3 On‐screen measurement tools 0.5 1
SE4 Setup verification tools 0.5 (0.5°) 1 (1°)
Annually
AE1 Independent quality control review Complete