. 2007 May 10;8(2):76–92. doi: 10.1120/jacmp.v8i2.2350

Table 2.

Tolerances a used in the validation of the analytical anisotropic algorithm

Test case TG‐53 situations Inner (%) Penumbra (mm) Outer (%) Buildup (%) TSF (%)
1. Open square fields Square fields 1.5 2 2 20 0.5
2. Extended SSD square fields SSD variations 1.5 2 2 40 1
3. Rectangular fields Max (rectangular, square fields) 2 2 2 20 0.5
4a. Wedged square fields Wedged 5 3 5 50 2
4b. Wedged, with 45‐degree collimator Wedged 5 3 5 50 2
5. Cerrobend‐blocked b mantle field Blocked 3 2 5 50 1
6. Isocentric, 10×10 cm at surface Square fields 1.5 2 2 20 0.5
7. Oblique incidence Max (SSD variations, external surface variations) 3 2 5 40 1
8. Asymmetric half‐beam Asymmetric 3 2 3 20 1
9. Oblique incidence, 45‐degree wedge Wedged 5 3 5 50 2
10. Multileaf collimator (MLC) MLC 3 3 5 20 1
a

The value of each tolerance was obtained from TG5320, published by the American Association of Physicists in Medicine.

b

Cerro Copper and Brass Company, Belfonte, PA.

TSF= total scatter factor (defined in section III.B); SSD= source‐to‐surface distance.