. 2007 May 10;8(2):76–92. doi: 10.1120/jacmp.v8i2.2350

Table 5.

Percentage differences a in the total scatter factor (TSF), by institution and energy, between measured values and those calculated by analytical anisotropic algorithm for each test case

TBCC CCI
Test case 6 MV (%) 15 MV (%) 6 MV (%) 15 MV (%)
1. Open square fields 0.4 1.3 0.3 1.0
0.6 0.6
0.5 0.4 0.6 0.7
2. Extended source‐to‐surface distance 0.6 1.5 0.9 1.1
0.5 0.5 0.5 0.3
3. Rectangular fields 0.3 0.8 0.6 0.9
0.4 0.8 0.0 0.6
4a. Wedged fields 1.4 1.0 1.4
0.6
1.2 1.0 0.5
0.5
1.1
3.0
0.5
0.6
4b. Open collimator, 45 degrees 0.3 0.1
Wedged collimator, 45 degrees 1.1 1.1 1.0
0.5
5. Mantle field 0.5
0.8
0.9 1.3
6. Isocentric 0.0
0.6
0.1 0.2
7. Oblique incidence
1.4
0.6
1.3
0.4
1.0
0.9
1.0
1.0
8. Asymmetric fields 0.3
0.2
1.4 0.3
1.0
2.3
9. Oblique, wedged
0.7
0.5 0.4
1.4
10. Multileaf collimator shaped 0.7 1.5 0.7 1.2
Mean 0.4 0.3 0.3 0.2
Range 1.4 1.5 1.4 1.3
1.4
3.0
1.3
2.3
a

Discrepancies greater than 2% are bolded.