Table 6.

Doses measured with a thermoluminescent dosimeter (TLD) in the thorax section of the Rando phantom a as compared with calculations by the Eclipse b analytical anisotropic algorithm (AAA)

AAA Pinnacle3
TLD location (Fig. 6) Measured dose c (Gy) Calculated dose (Gy) %Diff d Calculated dose (Gy) %Diff d
1
2.59±2.2
2.67
3.0
2.63
1.2
2
2.57±0.8
2.63
2.3
2.58
0.4
3
2.62±1.8
2.67
1.9
2.58 1.8
4
2.40±1.7
2.47
3.0
2.46
2.5
5
2.24±0.7
2.29
2.2
2.28
1.6
6
2.23±0.4
2.30
2.8
2.24
0.3
7
1.98±2.1
2.00
0.5
1.98 0.2
8
2.10±6.3
2.20
5.0
2.15
2.6
9
1.90±1.8
1.96
2.9
1.89 0.5
10
1.60±2.4
1.58
1.5
1.57 2.0
11
1.70±1.3
1.70
0.5
1.67 2.3
12
1.44±2.8
1.52
5.2
1.45
0.6
Mean difference
2.2
0.2
Maximum difference
5.2
2.6
a

The Phantom Laboratory, Salem, NY.

b

Varian Medical Systems, Palo Alto, CA.

c

Standard deviation of three measurements.

d

Differences are expressed as a percentage of the measured dose.