Skip to main content
. 2006 Aug 24;7(3):1–11. doi: 10.1120/jacmp.v7i3.2212

Figure 4. Comparison of peripheral dose distributions measured on a helical tomotherapy system and various conventional 6‐MV beams. The tomotherapy and 21‐EX measurements were taken with a 1×40cm2 field size. Mutic et al. (9) data measured at dmax with the MLC set to 15×15cm2. Stern (10) data measured at a depth of 5 cm with the MLC set to 10×10cm2. Giessen (11) averaged measurements for 4‐ to 25‐MV linear accelerators with a 5×5cm2 field size. TG‐36 values were measured at dmax for a field size of 15×15cm2.

Figure 4