. 2006 May 25;7(2):86–95. doi: 10.1120/jacmp.v7i2.2202

Table 1.

Ion chamber measured data (cGy/MU) at each field position with all eight fields irradiated used to calculate the monitor units (MU) required to generate the predetermined dose levels at each position on the XV film calibration exposure. All values were doubled for EDR film exposure.

Field 1 2 3 4 5 6 7 8 Dose (cGy)
detector position 1 0.8890 0.0015 0.0012 0.0002 0.0023 0.0012 0.0005 0.0001 16.3
2 0.8877 0.8749 0.0035 0.0007 0.0029 0.0020 0.0011 0.0004 32.1
3 0.8917 0.8881 0.8752 0.0025 0.0030 0.0029 0.0015 0.0006 48.2
4 0.8943 0.8900 0.8873 0.8734 0.0027 0.0021 0.0013 0.0008 64.1
5 0.0023 0.0012 0.0005 0.0001 0.8906 0.0015 0.0012 0.0002 80.3
6 0.0029 0.0020 0.0011 0.0004 0.8946 0.8803 0.0035 0.0007 95.7
7 0.0030 0.0029 0.0015 0.0006 0.8973 0.8937 0.8803 0.0025 112.0
8 0.0027 0.0021 0.0013 0.0008 0.8992 0.8962 0.8904 0.8762 128.1
MUs 18 18 18 18 90 17 18 18