Table 5.
Reactions to fields that fail to meet acceptance criteria, multiple responses accepted
Frequency of responses to the question “If the dose analysis for a beam (measured vs. planned) does NOT pass your standard criteria for passing, how do you respond?” | |
---|---|
Response | Response frequency (% of ) |
Examine the field(s) and TPS calculated dose to search for known limitations in the IMRT dose algorithm | 71.2% |
Examine the other beams in the plan to determine if the magnitude of the error might be significant | 61.2% |
Change the plan if necessary | 51.8% |
Examine the delivery records or devices (e.g. machine parameters, log files, physical devices, etc.) for errors | 42.4% |
Change the criteria to get a higher passing rate and record the justification | 39.6% |
Change the criteria to get a higher passing rate without recording any justification | 6.5% |
Not applicable, because my analyses always pass the criteria | 2.9% |
I do nothing | 0.0% |
I do something else | 59.0% |