Table 5.
Reactions to fields that fail to meet acceptance criteria, multiple responses accepted
| Frequency of responses to the question “If the dose analysis for a beam (measured vs. planned) does NOT pass your standard criteria for passing, how do you respond?” | |
|---|---|
| Response | Response frequency (% of ) |
| Examine the field(s) and TPS calculated dose to search for known limitations in the IMRT dose algorithm | 71.2% |
| Examine the other beams in the plan to determine if the magnitude of the error might be significant | 61.2% |
| Change the plan if necessary | 51.8% |
| Examine the delivery records or devices (e.g. machine parameters, log files, physical devices, etc.) for errors | 42.4% |
| Change the criteria to get a higher passing rate and record the justification | 39.6% |
| Change the criteria to get a higher passing rate without recording any justification | 6.5% |
| Not applicable, because my analyses always pass the criteria | 2.9% |
| I do nothing | 0.0% |
| I do something else | 59.0% |