. 2007 Sep 24;8(4):169–175. doi: 10.1120/jacmp.v8i4.2679

Table 5.

Effect of cadmium foil on the dose measured by TLD‐100 at 18 MV

Measurement location Dose (μGy/MU)
Distance a (cm) Depth b (cm) TLD‐700 TLD‐100
No Cd 0.25‐mm Cd 0.5‐mm Cd 0.75‐mm Cd
17 3.75 172.3 327.8 206.4 202.9 201.3
17 11.25 134.2 216.1 163.1 165.0 158.5
51.6 3.75 10.1 96.5 18.7 16.9 16.9
51.6 11.25 7.7 57.0 12.0 12.0 11.9
a

Distance from the central axis.

b

Depth from phantom surface.