. 2005 Nov 22;6(4):22–39. doi: 10.1120/jacmp.v6i4.2104

Table 4.

Disagreement (% error in central axis dose at dmax=1.7cm) between the measured and EPID‐calculated water phantom dose for 6 MV

Field size (cm2) Error (.%)
2×2
2.6%
5×5
0.9%
10×10
0.3%
20×20
1.0%
25×25
1.1%