Table 5.

Elekta Precise 60° wedge factors (WF) for 6‐MV and 10‐MV photons. Air and water wedge factors were measured with a 0.6 cm3 Farmer chamber. The difference (%) between WFair,155cm and WFEPID is reported in the last column.

Field size Energy In water In air In air iViewGT % Diff
(cm2)
(MV) 100 cm 100 cm 155 cm EPID
SSD+dmax
SCD SCD
5×5
6 0.286 0.288 0.287 0.248 15.7%
10×10
6 0.293 0.295 0.293 0.251 16.7%
20×20
6 0.306 0.307 0.304 0.260 16.9%
5×5
10 0.308 0.303 0.301 0.280 7.5%
10×10
10 0.311 0.310 0.308 0.282 9.2%
20×20
10 0.328 0.323 0.319 0.293 8.9%