Table 1.

Summary for lung cancer patients. Positive Dpt means portal image diaphragm is inferior to DRR (more like “inhale” than simulation). Spt is the SD of D for an individual patient. The estimated error in Dpt is Spt/(# films). Diaphragm excursions are estimated visually from a fluoroscopic movie acquired at simulation. The approximate period and duty cycle are estimated from the CT simulation reference session.

Patient #film Visual prompt
Dpt±estimated error (cm)
Spt (cm)
Range of D (cm) % within 0.5 cm of Dpt Estimated diaphragm excursion (cm) Estimated diaphragm excursion in gate (cm) Period(s) Duty cycle (%)
1_LU 9 no
0.14±0.09
0.27 (0.2, 0.6) 100%
~1.5
not available a 5 38
2_LU 12 no
0.05±0.11
0.38 (0.6, 0.4) 92% ~1.5 to 2.0
~0.5
5 29
3_LU 11 no
0.03±0.10
0.30 (0.6, 0.4) 91% ~1.0 to 1.5
~0.5
4 40
4_LU 9 no
0.04±0.16
0.48 (0.4, 1.2) 89%
~2.0
~0.5
4.5 27
5_LU 12 no
1.14±0.14
0.48 (1.6, 0.2) 92% ~2.5 to 3.0
~0.5
4 31
6_LU 12 yes
0.02±0.14
0.50 (1.0,0 .9) 67% ~1.5 to 2.0
~0.7
5 50
7_LU 12 yes
0.38±0.10
0.34 (10, 0.1) 92%
~2.0
~0.5
5 35
8_LU 12 no
1.12±0.09
0.32 (1.6, 0.7) 100%
~2.0
~0.5
5 39
9_LU 16 yes
0.35±0.10
0.40 (0.3, 1.1) 81% ~1.0–1.5
~0.5
5 36
10_LU 14 no
0.34±0.16
0.58 (0.4, 1.6) 50% no movie no movie 5 47
11_LU 10 no
0.13±0.13
0.40 (1.0, 0.4) 80% ~4.0–5.0 not available 4 42
12_LU 9 no
0.03±0.06
0.17 (0.4, 0.2) 100% no movie no movie 5 46
13_LU 17 yes
1.04±0.14
0.56 (0.01, 1.8) 65% 2.2
~0.8
4.4 38
14_LU 9 yes
1.41±0.17
0.51 (2.1, 0.3) 78% 1.5
~0.8
5 35
15_LU 12 yes
0.35±0.11
0.37 (0.2, 0.8) 83% 3.3
~1.0
6 39
16_LU 11 yes
0.19±0.09
0.30 (0.25, 0.55) 100% 2.3
~0.5
5.8 44
17_LU 10 no b
1.37±0.17
0.54 (1.75, 0.15) 90% 2.5
~0.7
5.2 42
18_LU 11 no b
0.05±0.17
0.57 (0.7, 1.6) 82% 1.5 atypical trace a 5 55
19_LU 14 yes
0.19±0.14
0.51 (0.5, 1.2) 71% 2.2
~1.0
4.7 51
20_LU 8 yes
0.25±0.14
0.4 (0.9, 0.4) 75% 1.5
~0.5
4.9 32
21_LU 11 yes
0.38±0.14
0.49 (0.35, 1.25) 73% 1.7 ~0.7 (atypical trace a ) 5.4 54
a

Fluoro movie connection to breathing trace corrupted; breathing trace for fluoro qualitatively different from CT and treatment.

b

Offered visual prompting but declined