Skip to main content
. 2004 Oct 21;5(3):37–54. doi: 10.1120/jacmp.v5i3.1982

Figure 9.

Figure 9

(a) Comparison between TPS calculated and measured relative dose for the uniform test field in the direction of leaf movement. Profile superposition shows an agreement better than 1% in the HD‐LG region. (b) Profile superposition in a perpendicular direction to leaf movement shows an agreement better than 1% in the HD‐LG region.