Skip to main content
. 2004 Oct 21;5(3):37–54. doi: 10.1120/jacmp.v5i3.1982

Figure 11.

Figure 11

(a) Comparison between TPS calculated and measured relative dose for the hole test field in the direction of leaf movement. Profile superposition shows an agreement better than 3% in the HD‐LG region. (b) Comparison in a direction perpendicular to leaf movement shows a dose difference similar to the uniform test field. Isodose superposition shows agreement better than 4% dose difference and 3‐mm distance to agreement.