Skip to main content
. 2005 Aug 17;6(3):133–142. doi: 10.1120/jacmp.v6i3.2110

Table 4.

Estimated dose uncertainties due to the finite grid size for dose calculations and measurements

Collimator (mm) Half of grid size (mm) Dose gradient (Gy/mm) Dose uncertainty (Gy)
8 0.6 18.1 10.9
14 1.0 9.1 9.1
18 1.25 7.3 9.1