Skip to main content
. 2005 Mar 17;6(1):101–113. doi: 10.1120/jacmp.v6i1.2020

Table 2.

Typical measured values of ESDs at different phantoms, values obtained from the mathematical model and typical α, β, and backscatter factors values

Examination Phantom ESD measured (mGy) ESD mathematical model (mGy) α(x1000) β BSF
AP abdomen ALD 3.00 2.29 9.849 1.988 1.35
PA chest CHE 0.20 0.25 2.580 2.016 1.47
AP chest CHE 0.36 0.32 6.893 2.016 1.35
LAT chest CHE 0.88 0.80 3.404 2.016 1.47
AP lumbar spine ALD 2.98 3.29 10.478 2.042 1.32
LAT lumbar spine ALD 7.41 8.11 10.241 2.040 1.35
LAT lumbo‐sacral joint ALD 8.30 7.42 12.256 2.062 1.35
AP pelvis ALD 2.66 2.84 9.627 2.084 1.32
PA skull SKU 1.91 1.73 9.204 2.006 1.29
LAT skull SKU 1.45 1.52 9.002 2.006 1.29
AP urinary tract ALD 2.50 2.79 10.066 2.006 1.35