Table I.

Neutron source strength (Q) values for various linac model and energy combinations.

Manufacturer Model Nominal MV Q values (×1012 neutrons per Gy) Std. Deviation (×1012) No. of Linac s Published Q values (×1012)
Varian 1800 10 0.06
Varian 1800 15 0.76
Varian 1800 18 1.22
Varian 2100C 18 0.96 0.11 17
Varian 2100C b 18 0.87 1
Varian 2300CD 18 0.95 0.03 2
Varian 2500 24 0.77 1
Siemens MD2 10 0.08 1
Siemens MD 15 0.20 0.02 2
Siemens KD 18 0.88 0.10 2
Siemens KD 20 0.92
Siemens Primus a 10 0.02 1
Siemens Primus a 15 0.12 1
Siemens Primus b 15 0.21 1
Siemens Primus 15 1 0.20
Elekta SL‐20 17 0.69
Elekta SL‐20 18 0.46 1
Elekta SL‐25 18 0.46 1
Elekta SL‐25 22 2.37
Elekta SL‐25 25 1.44 0.31 3
GE Saturne 41 12 0.24
GE Saturne 41 15 0.47
GE Saturne 43 18 1.32 1 1.50
GE Saturne 43 25 2.40
a

With MiMIC device in place.

b

With the multileaf collimator set to a 3.8cm2×3.8cm2 field.