Table III.

Allowable monitor units per week for floor below.

Lead (mm) Concrete (mm)       Noncontrolled
      Distance to occupied area (meters)
1 2 3 4 5
0 0 70 300 600 1000 1600
5 25 90 400 800 1400 2100
10 50 120 500 1000 1800 2800
15 75 150 600 1400 2400 3700
20 100 200 800 1800 3200 5000
25 125 300 1100 2400 4200 6500
1000 MU/min
Lead (mm) Concrete (mm)       Controlled
      Distance to occupied area (meters)
1 2 3 4 5
0 0 3 200 13 000 29 000 52 000 81 000
5 25 4 300 17 000 38 000 69 000 110 000
10 50 5 600 23 000 51 000 90 000 140 000
15 75 7 500 30 000 67 000 120 000 190 000
20 100 10 000 40 000 89 000 160 000 250 000
25 125 13 000 52 000 120 000 210 000 330 000
  1000 MU/min