Table 1.

Motion ranges, RMS errors, 3D normalized RMS errors and percentages of time for 3D errors exceeding 1, 2, 3 mm in simulation cases of ellipsoidal motions with noise added to the 2D marker positions.

Motion ranges (mm) RMS errors (mm) 3D normalized RMS errors % of time for 3D errors exceeding (mm)
x y z 3D 1 2 3
1.0 0.336 0.341 0.062 0.483 0.483 0.0 0.0 0.0
2.0 0.671 0.673 0.079 0.934 0.477 5.0 0.0 0.0
3.0 1.006 1.009 0.100 1.428 0.476 7.5 2.5 0.0
4.0 1.343 1.340 0.124 1.901 0.475 8.8 5.0 1.3
Average 0.839 0.841 0.091 1.187 0.478 5.3 1.9 0.3