Abstract
This paper reviews the current research on the speciation and determination of mercury by various analytical techniques, including the atomic absorption spectrometry (AAS), voltammetry, inductively coupled plasma optical emission spectrometry (ICP-OES), ICP-mass spectrometry (MS), atomic fluorescence spectrometry (AFS), spectrophotometry, spectrofluorometry, and high performance liquid chromatography (HPLC). Approximately 96 research papers on the speciation and determination of mercury by various analytical instruments published in international journals since 2015 were reviewed. All analytical parameters, including the limits of detection, linearity range, quality assurance and control, applicability, and interfering ions, evaluated in the reviewed articles were tabulated. In this review, we found a lack of information in speciation studies of mercury in recent years. Another important conclusion from this review was that there were few studies regarding the concentration of mercury in the atmosphere.
1. Introduction
Mercury is the only metal that exists in a liquid state among the elements in our modern periodic table. Determination and speciation studies of mercury attract researchers because of the toxicity of mercury to humans, as well as to other animals in the food web. The difference between the toxicity of mercury and that of other metals is that mercury easily accumulates in organisms. A few studies have reported bioaccumulation of mercury in various aquatic animals, such as fishes, pelagic seabirds, and earthworms [1–9].
This section describes the sources and fate of mercury in the environment and its toxicity.
1.1. Sources and Fate of Mercury in the Environment
Mercury can enter the environment from natural and/or anthropogenic sources. Natural sources of mercury include volcanoes, forest fires, cinnabar (ore), and fossil fuels, such as coal and petroleum. Anthropogenic sources are numerous; a large number of human activities are responsible for mercury deposition in the environment. Anthropogenic sources of mercury are landfills, dental preparations, and combustion processes, such as coal-fired power generation, medicinal waste incinerators, and municipal waste combustion. Manufacture of metals, alkali, and cement also releases mercury into the environment [10]. Anthropogenic sources are related to human activities in contaminated locations. This section describes the sources of mercury in the environment, reported from various parts of the world. Zhuang and Gao [11] reported higher concentrations of mercury in riverine sediments than in marine sediments and concluded that river transportation was the main source of mercury in southwestern Laizhou Bay, China. Kwon et al. [12] found that watershed runoff was the primary route of mercury transfer between lakes and forests.
Xu et al. [13] revealed that mercury concentration in soil has recently increased 3–10 times because of the combustion of fossil fuels combined with long-range atmospheric transportation processes. Shamsipur et al. [14] and Rajabi et al. [15] reported the determination of mercury in water samples using spectrometric and electrochemical techniques, respectively. Han et al. [16] found lower concentrations of wet-deposited mercury in forest areas of South Korea during summer because of precipitation. The concentration of mercury in the atmosphere was influenced by the seasons. In the atmosphere, coal combustion was the major source of gaseous elemental mercury, but traffic emissions contributed particulate mercury. Domestic pollutants are major sources of reactive gaseous mercury [17].
1.2. Toxicity and Health Implications of Mercury and Its Different Species
Researchers determine the concentration of mercury in environmental segments because of its toxic nature. Numerous journal articles have been published regarding the toxicity of mercury and its different forms. Yoshida et al. [18] reported on its neurobehavioral toxicity in mice exposed to low-level mercury vapor and methylmercury. Bucio et al. [19] studied the toxicity of mercury in a human hepatic cell line (WRL-68 cells). Results of this study indicated that higher doses of mercury cause cytotoxic effects with the release of lactate dehydrogenase from cells. Mercury exposure can cause neurodegeneration with oxidative stress in mitochondria [20]. Occupational exposure to mercury in workers of a fluorescent lamp factory in Egypt resulted in symptoms including emotional ability, memory changes, neuromuscular changes, and performance deficits in tests of cognitive function [21]. Mercury(II) and methylmercury toxicity can inhibit the human thioredoxin system. Mercury inhibition is selective for the thioredoxin system; mercury binds with selenol-thiol in the active sites of thioredoxin reductase [22]. Methylmercury reacts with the sulfhydryl groups throughout the human body and influences the functions of cellular and subcellular structures. Mercury toxicity in various forms can cause thyroid dysfunction because of the inhibition of 5′deiodonases, spermatogenesis because of accumulation in the testicles, and atrophy and capillary damage in thigh muscles [23]. Tonazzi et al. found a correlation between mitochondrial carnitine-acylcarnitine transporter inactivation and mercury toxicity in animals [24]. Mercury toxicity in humans can cause numerous neurological or psychiatric disorders not limited to autism spectrum disorders, Alzheimer's disease, Parkinson's disease, epilepsy, depression, and tremor. In rats, mercury(II) toxicity affects the central neurons and leads to cytoskeleton instability [25]. Exposure to organic forms of mercury, such as ethylmercury or methylmercury, can cause neurotoxic effects in developing mammals. Ethylmercury exposure in humans occurs because of immunization with thimerosal-containing vaccines [26]. The toxicity of mercury not only is limited to neurological effects in humans, but also causes vascular effects, such as increased oxidative stress and inflammation, thrombosis, endothelial dysfunction, dyslipidemia, and immune and mitochondrial dysfunctions [27]. Overall, the toxicity of mercury in animals and humans affects the cardiovascular, hematological, pulmonary, renal, immunological, neurological, endocrine, reproductive, and embryonic systems [28].
Plants are exposed to mercury compounds through the administration of antifungal agents. The toxicity of mercury affects seed germination, growth, and development in higher plants. It also causes the breakdown of photosynthesis by affecting chlorophyll and magnesium molecules [29]. Mercury toxicity induces oxidative stress in growing cucumber seedlings and results in plant injury [30]. Mercury that has accumulated in different forms within plants can cause phytotoxicity and impair numerous metabolic processes, including nutrient uptake, water status, and photosynthesis [31].
In this present study, we reviewed speciation and determination studies of mercury in different environmental samples using various analytical techniques, including the atomic absorption spectrometry (AAS), voltammetry, inductively coupled plasma optical emission spectrometry (ICP-OES), ICP-mass spectrometry (MS), atomic fluorescence spectrometry (AFS), spectrophotometry, spectrofluorometry, and high performance liquid chromatography (HPLC). Over 96 research papers published since 2015 in reputable international journals were reviewed. This review clearly summarizes the current research on speciation and determination studies of mercury from locations worldwide.
2. Reviews of the Determination of Mercury
The toxic nature of mercury and its different species encourage researchers to determine their concentrations in different environmental samples. Recently, a number of reviews were published concerning the determination of mercury, which described various factors regarding the concentrations of mercury in the environment. This section summarizes recent reviews of the determination of mercury.
Hanna et al. [32] reviewed the concentrations of mercury in freshwater fishes of Africa. They reviewed 30 identified studies in which the authors collected 407 Hg concentrations from 166 fish species, 10 types of invertebrates, and various plankton species from 12 countries in Africa. The authors concluded there was a lack of data regarding Hg concentrations in African countries. However, based on available data, Hg concentrations were lower than that of the World Health Organization (WHO) recommendations for commercially available fishes in Africa. Ferreira et al. [33] reviewed analytical strategies of sample preparation for the determination of mercury in food samples using a cold vapor atomic absorption spectrometry (CV-AAS), cold vapor atomic fluorescence spectrometry (CV-AFS), inductively couple plasma mass spectrometry (ICP-MS), voltammetry, and neutron activation analysis. Based on the reviewed papers, they concluded that the determination of mercury and its species in food samples with CV-AFS or CV-AAS was simpler and less expensive than other methods.
Colorimetric and visual assay determination of Hg(II) based on gold nanoparticles, fluorescent gold nanoparticles, gold nanorods, gold nanoflowers, and gold nanostars was reviewed by Chansuvarn et al. [34]. They reported that gold nanoparticles were the most promising luminescent nanomaterials for the detection of Hg(II) because of high selectivity and ultrasensitivity. Regarding analytical instruments, the UV-visible spectrophotometer was cost-effective for limited-budget laboratories for real-time monitoring of Hg(II) in environmental samples. Ariya et al. [35] reviewed physiochemical and biogeochemical transformation of mercury in the atmosphere and at atmospheric interfaces. The authors described the analytical methodology for speciation of mercury in the atmosphere, exchange of Hg between the atmosphere and aquatic interfaces, and exchange of Hg between the atmosphere and terrestrial environments. Shrivastava et al. [36] reviewed Hg detoxification mechanisms in plants. The authors found that Hg had harmful toxic effects on the molecular and physiobiochemical behavior of plants. Another important conclusion of this study was that most research was conducted on seed germination and shoot, root, and leaf morphology. Duarte et al. [37] reviewed the utility of disposable sensors for the detection of lead(II), cadmium(II), and mercury(II) in the environment. The paper describes analytical performance and the effect of certain factors, such as immobilization procedures and surface modification, on the analytical characteristics of the sensors. The authors found that disposable sensors used for single measurements of lead(II), cadmium(II), and mercury (II) in environmental samples had adequate intersensor reproducibility, sensitivity, and selectivity and very low detection limits. They concluded that the modified carbon paste electrode provided better determination of Hg(II) and As(III), because of superior deposition with linear and improved responses under the set of studied conditions. The authors stated that the disadvantages of using macroelectrodes included their expense and that they suffered from surface fouling even though they provided better sensitivity and selectivity for the determination of Hg(II) and As(III).
Jagtap and Maher [38] reviewed the measurement of mercury species in sediments and soils by HPLC coupled with ICP-MS. The authors recommended the extraction of Hg species for determination by distillation or use of 2-mercaptoethanol. They also recommended usage of C8 as the stationary phase and 2-mercaptoethanol as the mobile phase in HPLC for accurate quantification of methyl mercury in presence of large amounts of Hg(II). Gustin et al. [39] reviewed the measurement and modeling of mercury in the atmosphere. These authors reported that mercury in the atmosphere can exist in three different forms, gaseous elemental mercury (GEM), gaseous oxidized mercury (GOM), and particulate bound mercury (PBM). Among these forms, there was relative confidence in GEM measurements only, whereas the remaining two forms were less understood. These authors concluded that only through the comparison of multiple calibrated measurements could the results be determined accurately. McLagan et al. [40] reviewed passive air sampling of GEM in the atmosphere. They found that the performance of the passive air sampling method must be validated against active air monitoring systems with satisfactory precision and accuracy. Jackson and Punshon [41] reviewed recent advances in the measurements of arsenic, cadmium, and mercury in rice and other food materials. They described the challenges, state-of-the-art methods, and usage of spatially resolved techniques for arsenic and mercury within rice grains. However, this review focused mainly on the determination and speciation studies of arsenic rather than mercury. Duan and Zhan [42] reviewed recent use of nanomaterials-based (noble metal nanoparticles, fluorescent metal nanoclusters, semiconductors quantum dots, and carbon nanodots) optical sensors for Hg(II) detection. They concluded that the advantages of using nanomaterials for Hg(II) detection and removal included higher sensitivity and selectivity, simpler and more rapid procedures, and lower cost than that of conventional methods. Sun et al. [43] reviewed the recent progress in detection of Hg using surface enhanced Raman spectroscopy (SERS). They stated that substantial enhancement in detectable Raman signals coupled with a unique nanoparticle-based approach made SERS a powerful tool for the detection of Hg(II). Suvarapu and Baek [44] reviewed the speciation and determination of mercury using various analytical techniques. They discussed research papers published during 2013-2014 on these topics.
3. Discussion
In recent years, a large number of research articles were published regarding the determination and speciation of mercury using various analytical techniques. We have divided this section into four parts based on the analytical techniques used in the determination of type and levels of mercury. They are (i) spectrometric techniques (AAS, AFS, ICP-OES, MS, spectrophotometry, and spectrofluorometry), (ii) electrochemical techniques (voltammetry and potentiometry), and (iii) miscellaneous techniques.
The determination and speciation of mercury using spectrometric techniques, such as AAS, AFS, ICP-OES, ICP-MS, spectrophotometry, and spectrofluorometry are presented in Table 1. Table 2 represents the determination of mercury using electrochemical techniques, and Table 3 represents the determination of mercury using miscellaneous techniques. In these tables, we have incorporated all the analytical variables of merit, such as limits of detection, linearity range, quality control and assurance, applicability (analyzed samples), and interference reported in the determination of mercury.
Table 1.
Analytical parameters of reviewed research papers involving speciation and determination of mercury by spectrometric instruments (AAS, ICP-OES, ICP-MS, AFS, spectrophotometer, and spectrofluorometer).
| Analyte | Method | Supporting media | Analytical instrument | LOD | Linearity range | QA/QC studies | Analyzed samples | Interference study | Ref |
|---|---|---|---|---|---|---|---|---|---|
| Hg(II) | Fluorescence | Dithioacetal-substituted triphenylimidazole | Fluorescence spectrometer | 4.3 nM | 0–10 μM | Sensitivity and selectivity of the method evaluated | — | Cations, such as Ag+, CO2+, K+, Sn2+, Cu2+, Ni2+, Mn2+, Na+, Ca2+, Mg2+, Pb2+, Fe3+, and Cd2+ did not interfere in determination of Hg2+ | [96] |
|
| |||||||||
| Hg(II) | Fluorescence | 2-Aminoethyl piperazine and 4-chloro-7-nitrobenz-2-oxa-1,3-diazole | Fluorescence spectrometer | 0.57 μM | 0.00–16.6 μM | The method is selective over 18 metal ions. Recoveries of Hg(II) in water samples in the range of 95–98.2% | Water samples | Either no or a small fluorescence signal was observed for Na2+, K2+, Mg2+, Ca2+, Al3+, Ga3+, In3+, Cr3+, Mn2+, Fe2+, Fe3+, Co2+, Ni2+, Cu2+, Zn2+ Cd2+, and Pb2+ ions | [110] |
|
| |||||||||
| GEM | Acid digestion | Teflon filters | ICP-MS | 0.015 M | — | SRM 1633c was analyzed. Recoveries were in the range of 84–112% | Particulate matter | — | [45] |
|
| |||||||||
| Hg(II) | CVG | LiAlH4, SnCl2/THB | HG-AFS | 0.0004 μM | The RSD values were less than 7.0% for 11 measurements. CRMs were analyzed | Soil, water, and human hair samples | 10 mg L−1 for Fe3+, 20 mg L−1 for Co2+, 10 mg L−1 for Ni2+, and 20 mg L−1 for Cu2+. For arsenic and antimony, no interference from 25 mg L−1 Cu2+, Co2+, or Ni2+ was observed | [89] | |
|
| |||||||||
| Hg(II) | Photochemical vapor generation | — | ICP-OES | 0.006 μM | Up to 0.25 μM | Recoveries of Hg(II) in reals samples were in the range of 79–112% | Petroleum production water | K+, Ba2+, Sr2+, Mg2+, Cu2+, Pb2+, and Zn2+ were over the concentration ranges studied, whereas the presence of Ca2+, Ni2+, Sb3+, As3+, Se4+, Fe3+, and Cr3+ was significant | [111] |
|
| |||||||||
| Hg(II) | Acid digestion | — | ICP-MS | — | 0.0005–0.5 μM | Precision of the method for Hg2+ is 5.6% with six replicates | Fish samples | Hg(II) was analyzed along with Cd2+ and Pb2+ | [112] |
|
| |||||||||
| Total Hg | Wet digestion | — | AAS | — | — | — | Fish samples | Hg(II) was analyzed along with Cd2+ and Pb2+ | [68] |
|
| |||||||||
| Hg speciation | Acid digestion | — | ICP-MS | 0.00004 μl/L | — | NIST 612 was analyzed | Petroleum hydrocarbons | — | [88] |
|
| |||||||||
| Total Hg | Chemical vapor generation | Nonionic surfactants | AFS | 22.5 μM | — | CRMs (GBW08603, GBW(E) 080401 and GBW(E)080402) were analyzed | Natural water samples | Severe interference of cations was observed at 10 mg L−1 concentration | [58] |
|
| |||||||||
| Total Hg | Sequential extraction | — | CV-AAS | — | — | Comparison of the results with other methods was performed | Thar coal | — | [102] |
|
| |||||||||
| Hg(II) | Aptasensor | Core-shell Ag@SiO2 nanoparticles | AFS | 0.33 nM | 0–1.2, 1.2–14 nM | Recoveries were over 94% for the determination of Hg(II) in real samples. The RSD values for Hg(II) determinations were lower than 5.1% | Real water samples | Selective in presence of Cd2+, Pb2+, Zn2+, Mn2+, Co2+, Fe2+, Cu2+, and Ag+ | [113] |
|
| |||||||||
| Hg(II) | Chemosensor | Porphyrin-thymine conjugates | Luminescence spectrometer | 6.7 nM | — | Method reversibility was reported | — | Interference of Zn2+, Cu2+, Ni2+, Mn2+, Mg2+, Pb2+, and Cd2+ was inhibited | [97] |
|
| |||||||||
| Hg | PVG and CVG | NaBH4/SnCl2 | PVG-AAS, CV-AAS (NaBH4) and CV-AAS (SnCl2) | 0.0006, 0.0005, and 0.0002 μM | 0.012–0.05 μM | The accuracy was evaluated by assessing recoveries in spiked samples which were in the range of 84–108.3% | Glycerin samples | — | [91] |
|
| |||||||||
| Total Hg | Solid sampling analysis | — | GF-AAS | 0.0014 μl/L | — | SRM (NIST-695) was analyzed. The RSD values were better than 8.2% for five replicates | Phosphate fertilizers | — | [53] |
|
| |||||||||
| DGM, TGM | — | — | CV-AFS | 1.35 × 10−8 μM (DGM), 1.35 × 10−8 μM (TGM) | — | Air-sea interface of Minamata | — | [46] | |
|
| |||||||||
| Methyl Hg | Distillation and solvent extraction | KBr/CuSO4 | GC-ICP-MS | 2.0 × 10−5 μg g−1 | — | CRM of ERM-CC580 was analyzed | Peat soil | — | [59] |
|
| |||||||||
| Total Hg | Ultrasound extraction | NaBH4/isoamyl alcohol, thiourea | CV-AAS | 70 μM | — | The RSD values of Hg determinations in vinegar was less than 8.11% | Vinegar | Vanadium also determined in the same samples | [108] |
|
| |||||||||
| Total Hg | Slurry sampling | CV-AAS | 150 μM | — | The RSD values in the mercury determinations was less than 10.89%. The recoveries were in the range of 85–106% | Iron supplement | — | [103] | |
|
| |||||||||
| Total Hg | NaBH4 | AES | 0.00004 μg g−1 | — | — | Sea food | — | [69] | |
|
| |||||||||
| Hg(II) | Preconcentration | Metal-organic frame work | CV-AAS | 0.05 μM | — | SRMs (DOLT-4 and DORM-2) were analyzed. The RSD values in the determination of Hg(II) was less than 10% | Sea food samples | Majority of cations did not interfere in the determination of Hg(II) at pH 6.25 experimental condition | [54] |
|
| |||||||||
| Total Hg | Solid sampling | — | HR-AAS | 2.0 × 10−5 μg (sediment), 9.6 × 10−5 μg (marine biota) | 2.0 × 10−5–0.004 μg (sediment), 2.0 × 10−5–0.025 μg (marine biota) | CRMs (PACS-2, IAEA-405, SRM 2703, BCR-464, IAEA-436, DORM-2, MA-ROPME-2/TM) were analyzed | Sediment and marine biota samples | — | [55] |
|
| |||||||||
| Total Hg | Thermal desorption | — | AAS | 0.0006 μg g−1 | — | CRMs (CRM-1515, MESS-3 and TORT-2) were analyzed with recoveries 96.0–104.8% | Fish and sea food samples | — | [60] |
|
| |||||||||
| Total Hg | ISO guide 34 | — | CV-ICP-MS | 8.0 × 10−5 μg/g | 1.9–50 × 10−5 μg/g | CRM (BCR-579) was analyzed | Sea water | By using cold vapor generation spectral interferences were avoided | [61] |
|
| |||||||||
| Total Hg | Acid digestion | Nitric and perchloric acid | AAS | 0.0004749 μg/g | 0.0002–0.01 μM | The mean recovery of Hg was 78.65%. RSD values for interday precision of Hg was 7.17% | Cream cosmetics |
Along with Hg, zinc was also determined | [109] |
|
| |||||||||
| Total Hg | Extraction | NaBH4 | CV-AAS | 0.003 μM | 0.05–0.5 μM | The recoveries of Hg in oil samples were in the range of 80–103% | Oil samples | — |
[104] |
|
| |||||||||
| Total Hg | Thermal desorption | — | TDA-AAS | 0.025 μg/g (LOQ) |
— | Mean recovery of Hg in real samples was 94.2% and SD was 3.5% | Sea food | — | [105] |
|
| |||||||||
| Total Hg | Method EPA 7473 | Teflon | CV-AFS | 0.0006 μg g−1 | 0.002–0.08 μg·g−1 | CRM (BCR-279) was analyzed and RSD in the determination of Hg in seaweeds is less than 10% | Seaweeds | — | [62] |
|
| |||||||||
| Hg(II) | Immobilization | Dithizone | FAAS | 2.0 × 10−9 M | 1.1 × 10−8–2.0 × 10−6 M | The coefficients of variation for Hg(II) was found to be 2.7% | Industrial wastewater, spiked tap water, and natural water |
Except Cu2+, the other ions (Mn2+, Ni2+, Pb2+, Co2+, Cd2+, Fe2+, Fe3+, and Al3+) did not interfere up to 50-fold excess | [114] |
|
| |||||||||
| MeHg | Online preconcentration | — | HPLC-CV-AFS | 40000 μM | 5–2500 μM | CRMs (NIES CRM no. 13 and IAEA-085) were analyzed. Recoveries of MeHg from real samples were in the range of 91.4–101.8% | Sewage, river, and seawater samples | — | [47] |
|
| |||||||||
| Hg(II) | Solid phase extraction | Ion imprinted polymeric nanomaterials | CV-AAS | 0.18 μM | — | The RSD values for eight replicates was 4.2% | Water and human hair samples | In presence of large amounts of Cu2+, Ni2+, Cd2+, Zn2+, Mn2+, Pb2+, Fe3+, and Cr3+ ion Hg(II) was effectively determined | [115] |
|
| |||||||||
| Hg(II), MeHg | Rapid extraction | — | HPLC-ICPMS | 0.0002 (Hg2+), 0.0001 (MeHg) μg g−1 | — | CRMs (TORT-2 and DORM-2) were analyzed | Fish samples | — | [63] |
|
| |||||||||
| Speciation | Liquid-liquid microextraction | Ionic liquid vortex-assisted | HPLC-CV-AFS | 3.4–6.1 × 10−6 μg/g | 0.0001–0.07 μg/g | The RSD values were less than 6.4% | Sediment samples | — | [92] |
|
| |||||||||
| Total Hg | Solid sampling | — | TDA-AAS | 0.001 μg g−1 | 0.025–0.2 μg g−1 | Recoveries of Hg from real samples were in the range of 89–99% | Fish and shrimp samples | — | [106] |
|
| |||||||||
| Hg speciation | Cloud point extraction | Polyethylene glycol | Spectrophotometer | 0.045 μM | 0.05–0.5 μM | The RSD values of the method were below 2.6% | River water and river sediment | No interference of Cd2+, Bi3+, and Pb2+ was observed in the determination of Hg(II) | [107] |
|
| |||||||||
| Hg(II) | Colorimetric | 1,5-diphenylthiocarbazone | Flow injection spectrophotometer | 0.15 μM | 0.25–7.5 μM | The results were compared with the data obtained with ICP-MS | Cosmetics and Thai traditional medicines | 20-fold Fe2+, Zn2+, and Cu2+, 40-fold Pb2+, 50-fold Al3+, Fe3+, and Mn2+ did not interfere in the determination of Hg(II) | [94] |
|
| |||||||||
| Hg(II) | Colorimetric, fluorescence | Calixpyrrole hydrazide | Spectrofluorometer | 1 nM | 1 nM–1 μM | — | Groundwater and industrial effluent water | No interference of Pb(II), Cd(II), Mn(II), Fe(III), Ni(II), Zn(II), Hg(II), Co(II), and Cu(II) was observed | [70] |
|
| |||||||||
| Hg(II) | Fluorescence probe | Chitosan hydrogel | Fluorescence spectrophotometer | 0.9 nM | 5.0–50 nM | — | Water samples | Cations such as Fe3+, Co2+, Pb2+, Cu2+, Cd2+, Ni2+, and Zn2+ did not interfere in the determination of Hg(II) | [71] |
|
| |||||||||
| Hg(II) | Fluorescence sensor | CdTe quantum dots | Fluorescence spectrophotometer | 4.0 nM | 6.0–450 nM | RSD values were less than 4.15% | Lake water samples | Interference of 10-fold Pb2+, Cu2+, and Ag+ was less than 7% | [116] |
|
| |||||||||
| Hg(II) | Fluorescent chemosensor | DA | Fluorescence spectrophotometer | 0.0028 μl/L | — | Theoretical and experimental results were in good agreement with each other | — | Simultaneous determination of Ag+ and Cu2+ was reported | [98] |
|
| |||||||||
| Hg(II) | Time-gated fluorescent sensing | Thymine | Spectrofluorometer | 0.11 nM | 0.20–10 nM | Recoveries of Hg(II) in environmental water samples were in the range of 93.75–102.5% | Drinking water samples | No interference of Ag+, Co2+, Ni2+, Ca2+, Cd2+, Al3+, Fe3+, Au3+, Cr2+, Mn2+, Pb2+, Cu2+, Mg2+, Zn2+, and Ba2+ was reported | [117] |
|
| |||||||||
| Hg(II) | Colorimetric | Gold nanoparticles | Spectrophotometer | 0.5 nM | 0.5–300 nM | CRM (GBW (E) 080392) was analyzed and the recoveries were found in the range of 88.9–106% | Tap water and lake water | Cr3+, Mn2+, Co2+, Ni2+, Cu2+, Fe3+, Zn2+, Cd2+, and Pb2+ ions did not interfere in the determination of Hg2+ | [64] |
|
| |||||||||
| Hg(II) | Fluorescence | Schiff base | Fluorescence spectrophotometer | 2.82 × 10−6 M | — | Effectiveness of the method was proved by confocal fluorescence microscope | Living cells | — | [95] |
|
| |||||||||
| Hg(II) | Colorimetric | Silver nanoparticles | Spectrophotometer | 1.18 × 10−9 M | 10–50 nM | A good linear correlation (R2 = 0.9799) was obtained for different concentrations of Hg(II) and absorbance ratio | Lake, seawater, and groundwater | Fe2+, Fe3+, Cr6+, Pb2+, Mn2+, Al3+, Ni2+, Cr3+, Cd2+, Mg2+, and Zn2+ did not interfere up to 1000 times of detection limit of Hg(II) | [118] |
|
| |||||||||
| Hg(II) | Chemosensor | Dimeric binol-based chemosensor | Spectrofluorometer | 4.4 × 10−7 M | — | — | Wastewater samples | 100 equivalents of Na+, K+, Mg2+, Ce3+, Ca2+, Ba2+, Cd2+, Mn2+, Co2+, Ni2+, Cu2+, Cr3+, Zn2+, Pb2+, Fe2+, Fe3+, Al3+, and Ag+ | [72] |
|
| |||||||||
| Hg(II) | Colorimetric | Bovine serum albumin | Spectrophotometer | 7.2 nM | 0–120 nM | Results were compared with ICP-MS | Drinking water samples | No interference of Na+, Mg2+, Ca2+, Cd2+, Mn2+, Co2+, Ni2+, Cu2+, Zn2+, Pb2+, Fe3+, and Au3+ was observed | [119] |
|
| |||||||||
| Hg(II) | Fluorescence | Gold nanocluster | Spectrofluorometer | 30 nM | — | Recoveries of Hg(II) in spiked samples were in the range of 97.7–99.3% | Lake water samples | No interference of Na+, Mg2+, Ca2+, Ni2+, Cu2+, Zn2+, Ce3+, Pt4+, and Al3+ was observed | [120] |
|
| |||||||||
| Hg(II) | Colorimetric | Rhodamine B | Spectrofluorometer | 1.71 × 10−6 M | — | — | Spiked tap water samples | Interference of several ions was negligible in the determination of Hg(II) | [73] |
|
| |||||||||
| Hg(II) | Fluorescence | — | Fluorescence spectrometer | 9.56 × 10−9 M |
— | — | — | Hg(II) can be detectable in presence of Fe3+, Cu2+, Co2+, Ni2+, Cd2+, Pb2+, Zn2+, and Cr3+ | [74] |
|
| |||||||||
| Hg(II) | Adsorption | Rhodamine | Fluorescence spectrophotometer | 3.42 × 10−6 M |
0–6.0 μM | — | Drinking and lake water | No interference of Cd2+, Co2+, Cu2+, Fe3+, Mn2+, Ni2+, Pb2+, and Zn2+ at 581 nm | [75] |
|
| |||||||||
| Hg(II) | Fluorimetric | Coumarinyldithiolane | Fluorescence spectrophotometer | — | 0.06–1.5 μM | — | Aqueous solutions | No influence of Al3+, Zn2+, Co2+, Ni2+, Cu2+, Cd2+, Cr3+, and Pb2+ on the determination of Hg(II) in presence of probe | [76] |
|
| |||||||||
| Hg(II) | Fluorescence sensors | Peanut shell | Fluorescence spectrometer | 8.5 × 10−9 M | 0–19 × 10−8 M | — | Lake water | The method was selective for Hg(II) | [77] |
|
| |||||||||
| Hg(II) | Colorimetric | L-Arginine | Spectrophotometer | 5 nM | 1–20 and 20–600 μM | — | Food samples | No interference from Cd2+, Co2+, Cu2+, Ni2+, and Pb2+ was observed | [78] |
LLME: Liquid-liquid microextraction; CRM: certified reference material; CVG: chemical vapor generation; THB: tetrahydroborate; RSD: relative standard deviation; LOQ: limit of quantification; MIP-OES: microwave-induced plasma optical emission spectrometer; PPT: poly(1,4-bis-(8-(4-phenylthiazole-2-thiol)-octyloxy)-benzene); LSPR: localized surface plasmon resonance; DA: dimethylaminocinnamaldehyde-aminothiourea. Analytical Instruments. CV-AAS: Cloud Vapor Atomic Absorption Spectrometer; GF-AAS: Graphite Furnace AAS; ICP-OES: Inductively Coupled Plasma Optical Emission Spectrometer; ICP-MS: ICP-Mass Spectrometer; ICP-AES: ICP-Atomic Emission Spectrometer; HPLC: High Performance Liquid Chromatography; AFS: Atomic Fluorescence Spectrometer.
Table 2.
Analytical parameters of reviewed research papers involving speciation and determination of mercury by electrochemical instruments.
| Analyte | Method | Supporting media | Analytical instrument | LOD | Linearity range | QA/QC studies | Analyzed samples | Interference study | Ref |
|---|---|---|---|---|---|---|---|---|---|
| Hg(II) | Biosensor | Y-shaped DNA | Square wave voltammeter | 0.094 nM | 1 nM–5 μM | Selectivity, sensitivity, and repeatability were studied | River water samples | Interferences of Cu2+, Al3+, Co2+, Fe3+, Zn2+, Ni2+, Cd2+, Ba2+, Cr3+, Mg2+, and Pb2+ were reported | [121] |
|
| |||||||||
| Hg(II) | Preconcentration | N-Octylpyridinium | Stripping voltammeter | 0.0015 μM | 0–0.5 μM | The RSD of the method was 10% | Tap, pond, and wastewaters | No significant interference of 100 μg L−1 of Cu2+, Pb2+, Cd2+, and Zn2+ was observed in the determination of Hg2+ | [122] |
|
| |||||||||
| Hg(II) | Electrochemical | Screen printed carbon electrode | Anodic stripping voltammeter | 0.005 μM | 0.005–0.5 μM | Accuracy of the method was evaluated with ICP/MS | Groundwater | Interference of Cu2+, Co2+, Fe2+, Zn2+, Ni2+, Cd2+, Mn2+, Mg2+, and Pb2+ was negligible in the determination of Hg2+ | [123] |
|
| |||||||||
| Hg(II) | Electrochemical sensor | 1-(2, 4-Dinitrophenyl)-dodecanoyl thiourea | Cyclic, square wave and differential pulse voltammeter | 0.0032 μM | Up to 0.01 μM | The RSD of the method was 3.5% | Drinking and tap water samples | 5-fold Cu2+, Cd2+, Pb2+, and Zn2+ did not interfere in the determination of Hg(II) | [124] |
|
| |||||||||
| Hg(II) | Electrochemical | N-PC-Au | Anodic stripping voltammeter | 0.35 nM | 0.001–1 μM | — | Drinking water | The electrode was not affected by the presence of Zn2+, Pb2+, Cu2+, and Cd2+ ions in the determination of Hg(II) | [79] |
|
| |||||||||
| Hg(II) | Electrochemical sensor | Modified gold nanoparticles | Cyclic voltammeter | 7.5 μM | 5.0–50 μM | — | Spiked water samples | The method is selective towards the presence of Zn2+, Cd2+, Pb2+, Cu2+, Ni2+, and Co2+ ions | [80] |
|
| |||||||||
| Hg(II) | Electrochemical | N-doped graphene electrode | Differential pulse voltammeter | 0.05 μM | 0.2–9 μM | The RSD of Hg determination with six repetitions was 2.1% | Simultaneously Cd2+, Cu2+, and Pb2+ were determined along with Hg2+ | [99] | |
|
| |||||||||
| Hg(II) | Electrochemical sensor | Screen printed carbon electrode | Differential pulse anodic stripping voltammeter | 0.0001 μM | 0.0002–0.01 μM | Recovery of Hg(II) was found as 106% | Real water samples | High tolerance limits were observed for Fe3+, Zn2+, and Cd2+ but lower tolerance limits for Pb2+ and Cu2+ were found | [125] |
|
| |||||||||
| Hg(II) | Electrochemical sensor | DNA probe | Cyclic and square wave voltammeter | 5.6 nM | 10–100 nM | — | — | 10-fold Pb2+, Mn2+, Zn2+, Ni2+, Cu2+, Fe2+, Ba2+, and Cd2+ did not interfere in the determination of Hg(II) | [81] |
|
| |||||||||
| Hg(II) | Electrochemical | Carbon ionic liquid paste electrode | Anodic stripping voltammeter | 0.1 nM | 0.5–10 nM and 0.08–2 μM | — | Wastewater samples | Over 30-fold Zn2+, Cr3+, and Pb2+ and over 45-fold Cd2+, Cu2+, Ni2+, and Mn2+ interfered in the determination of Hg(II) | [82] |
|
| |||||||||
| Hg(II) | Electrochemical | Carbon paste sensor | Potentiometer | 1.95 × 10−9 M | 4.00 × 10−9–1.30 × 10−3 M | Reproducibility of the method was reported | Water samples | Selective coefficients of various cations for Hg(II) selective sensors were reported | [126] |
|
| |||||||||
| Hg(II) | Biosensor | Thymine | Differential pulse and cyclic voltammeter | 0.08 nM | 0.5–5000 nM | Recoveries of Hg(II) in real samples were in the range of 96.4–103% | Water and human serum | Selective in presence of Al3+, Ba2+, Cd2+, Co2+, Cr3+, Fe3+, Mn2+, Pb2+, and Zn2+ | [127] |
|
| |||||||||
| Hg(II) | Biosensor | Cyclic dithiothreitol | Cyclic voltammeter | 28 pM | 0.1 nM–5 μM | Recoveries of Hg(II) in water samples were in the range of 98.8–104% | River water samples | Excellent selectivity for Hg(II) detection was observed in presence of Cd2+, Pd2+, and Co2+ | [128] |
|
| |||||||||
| Hg(II) | Biosensor | Methylene blue | Cyclic voltammeter | 8.7 × 10−11 M | 1.0 × 10−10–5.0 × 10−7 M | The RSD of the sensor was 5.25% for 10 replicates indicating the good reproducibility | Tap and river water samples | Cd2+, Ba2+, Pb2+, Ni2+, Cu2+, Zn2+, Mn2+, Ca2+, Co2+, Mg2+, and Ag+ did not interfere up to 250 nM in presence of 50 nM of Hg(II) | [129] |
|
| |||||||||
| Hg(II) | Electrochemical | PVC membrane sensor | Potentiometer | 3.2 × 10−9 M | 1.0 × 10−8–5.0 × 10−3 M | RSD values for synthetic samples measurements were less than 3.10% | Wastewater samples | The selectivity coefficients for various ions were in the range of 1.0 × 10−4–4.5 × 10−4 M | [130] |
|
| |||||||||
| Hg(II) | Electrochemical | Copper film electrode | Anodic stripping voltammeter | 0.0005 μM | 0.05–0.5 μM | The RSD value for 12 replicates of Hg determination was 4.5% | — | Simultaneously mercury and lead are determined | [100] |
|
| |||||||||
| Hg(II) | Electrochemical | Carbon nanotubes | Anodic stripping voltammeter | 0.025 μM | 0.1–100 μM | The RSD value for six replicates was 1.93% | River and industrial wastewater | Up to 200-fold Pb2+, Cu2+, Cd2+, Zn2+, Ni2+, and Mn2+ did not interfere in the determination of Hg(II) | [131] |
|
| |||||||||
| Hg(II) | Electrochemical sensor | Mesoporous carbon nanofibre | Anodic stripping voltammeter | 0.3 nM | 5–500 nM | The RSD values in the determination of Hg(II) in real samples were less than 2.3% | Yellow river, China | The proposed electrode avoids the interferences of Cd2+, Pb2+, and Cu2+ | [132] |
|
| |||||||||
| Hg(II) | Potentiometric sensor | MWCNTs | Potentiometer | 3.1 × 10−9 M | 4.0 × 10−9–2.2 × 10−3 M | The recoveries of Hg(II) were in the range of 99–102% | Aqueous samples | The proposed method was highly selective towards the determination of Hg(II) in presence of some other interfering ions in aqueous samples | [133] |
|
| |||||||||
| Hg(II) | Electrochemical | Rotating silver electrode | Square wave voltammeter | 4.61 × 10−8 M | 1.0 × 10−7–8.0 × 10−4 M | The RSD for seven replicates was 2.19% | Milk and breast milk | No interferences of copper, cobalt, iron, and zinc were observed | [134] |
|
| |||||||||
| Hg(II) | Electrochemical | Graphene modified with silver | Differential pulse voltammeter | 3.38 × 10−8 M | 5.0 × 10−8–1.0 × 10−4 M | The RSD for eight replicates was 2.25% | Leachate samples | Even 200 times excess of Al3+, Cd2+, Co2+, Ni2+, Pb2+, Fe2+, Fe3+, and Zn2+ did not interfere | [135] |
|
| |||||||||
| Hg(II) | Electrochemical | Graphene oxide | Cyclic voltammeter | 0.035 nM | 0.1–100 nM | The RSD value in the reproducible test was 4.5% | River water samples | Even 10 times higher concentrations of Co2+, Mn2+, Pb2+, and Fe3+ did not interfere in the determination of Hg(II) | [136] |
|
| |||||||||
| Hg(II) | Electrochemical | Gold nanoparticles | Differential pulse anodic stripping voltammeter | 0.0001 μM | 0.0005–0.05 μM | Recoveries of Hg(II) in real samples were in the range of 87–102% | Tap and lake waters, milk, and soils | 1000-fold Zn2+, Cd2+, Pb2+, Mn2+, Co2+, and Cu2+ did not interfere in the determination of Hg(II) | [137] |
|
| |||||||||
| Hg(II) | Electrochemical | Gold nanoparticles | Stripping voltammeter | 1 μM | — | Water samples | — | [83] | |
|
| |||||||||
| Hg(II) | Electrochemical | Graphene-Au modified electrode | Square wave voltammeter | 0.001 aM | 1.0 aM–100 nM | The RSD values for triplicate measurements was less than 4.46% | Spiked tap and river waters and landfill leachate samples | Even 500 nM of Cd2+, Co2+, Cr2+, Cu2+, Mn2+, Ni2+, Pb2+, Zn2+, Al3+, and Fe3+ did not interfere in the determination of 10 nM of Hg(II) | [48] |
|
| |||||||||
| Hg(II) | Electrochemical | Graphene/CeO2 | Differential pulse anodic stripping voltammeter | 2.187 × 10−11 M | 0.002–0.12 μM | — | Wastewaters | Simultaneously Cd2+, Pb2+, Cu2+, and Hg2+ were determined | [84] |
|
| |||||||||
| Hg(II) | Electrochemical | Graphene quantum dots | Anodic stripping voltammeter | 0.02 nM | 0.02–1.5 nM | Recoveries from spiked samples were in the range of 96.6–101% | Spiked samples | Cu2+ was also determined along with Hg(II) | [138] |
|
| |||||||||
| Total Hg | Liquid-liquid microextraction | Screen printed carbon electrodes | Square wave anodic stripping voltammeter | 0.00005 μM | 0.0025–0.05 μM | The recoveries in the determination of mercury in real samples were in the range of 95–108% | Tap, river, and bottled and industrial wastewaters | — | [139] |
|
| |||||||||
| Total Hg | Electrochemical sensing | Zinc oxide quantum dots | Linear sweep voltammeter | 0.005 μl/L | 0.005–0.05 μl/L | — | River and groundwater | Except Cd2+, the other ions, such as Zn2+, Pb2+, and As3+ did not interfere | [85] |
|
| |||||||||
| Total Hg | Electrochemical | Gold nanoparticles | Quartz crystal microbalance | 0.15 nM | 3–300 nM | The results were compared with CV-AAS technique. The RSD was found to be less than 7% | Water and sediment samples | Interference of Cu2+, Cr3+, Pb2+, and Cd2+ was reported | [140] |
|
| |||||||||
| Hg(0) | Electrochemical | Gold-based microsensor | Quartz crystal microbalance | — | — | The results were accurate and within 8% of the concentrations reported by EPA certified samples | Industrial flue gas | — | [141] |
|
| |||||||||
| Hg(0) | Electromechanical | — | Quartz crystal microbalance | 2.42 × 10−8 μM | — | Selectivity of the instruments for mercury was 84% | — | — | [142] |
|
| |||||||||
| Hg(0) | Electrochemical | Silver/gold core/shell nanowire monolayer | Quartz crystal microbalance | 0.039 μM | — | Repeatability of the results was always greater than 87% | Industrial gas effluents | — | [143] |
N-PC-Au: nitrogen-doped porous carbon-gold nanocomposite; MWCNTs: multiwalled carbon nanotubes.
Table 3.
Analytical parameters of reviewed research papers involving speciation and determination of mercury by miscellaneous techniques.
| Analyte | Method | Supporting media | Analytical instrument | LOD | Linearity range | QA/QC studies | Analyzed samples | Interference study | Ref |
|---|---|---|---|---|---|---|---|---|---|
| Hg(II) | Colorimetric | Gold nanoparticles | Dark-field microscope | 1.4 pM | — | Recoveries were 98.3 and 110.0% for river and industrial wastewater, respectively | River and industrial wastewater | 25 nM concentrations of Pb2+, Ni2+, Fe2+, Cd2+, Zn2+, Co2+, and Mn2+ did not interfere in the determination of Hg (II) | [49] |
|
| |||||||||
| Hg(II) | Electrochemiluminescence | Gold nanoparticles | Potentiostat PG340 | 5.1 pM | — | Results were compared with AFS measurements | Tap and lake waters | The method was selective for Hg determination in presence of Cd2+, Co2+, Cu2+, Fe2+, Mg2+, Mn2+, Pb2+, Al3+, and Fe3+ | [50] |
|
| |||||||||
| Hg(0) | Thermal desorption | — | Direct milestone analyzer | — | — | Accuracy was verified with testing the SRM (NIST-2711) and CRM (GBW-GBW 08301 RCV 8221) | Soil samples | — | [56] |
|
| |||||||||
| Hg(II) | Electrochemiluminescence | γ-Polyglutamic acid-grapheme-luminol | Chemiluminescence analyzer | 1.0 × 10−6 μl/L | 2.0 × 10−6–0.02 μl/L | The RSD values for reproducibility of biosensor were 6.2%; the results were compared with ICP-MS | River water samples | No interference of Pb2+, Zn2+, Cu2+, Mg2+, and Cd2+ was observed | [144] |
|
| |||||||||
| Hg speciation | Liquid-liquid-liquid microextraction | 18-crown-6 | Electrophoresis | 0.005–0.03 (Hg2+), 0.004–0.027 (Me Hg), 0.001–0.0075 (PhHg) μM | 0.01–1 μM | The RSD values of the reproducibility tests were less than 13.0% | Hair and water samples | — | [90] |
|
| |||||||||
| Hg(0) | UV-light generation | Multimode diode lasers | Photomultiplier modules | 0.12 μM | 0–60 μM | The coefficient of linear regression was obtained as R2 = 0.998 | — | — | [101] |
|
| |||||||||
| Total Hg | — | — | Direct mercury analyzer | — | — | — | Fish samples | Vanadium also determined along with mercury | [86] |
|
| |||||||||
| MeHg | ISO-17025 | — | Advanced mercury analyzer | 9.0 × 10−6 μg | 9.0 × 10−6–0.003 μg | CRMs (IAEA-436, DOLT-2, TORT-2, IAEA-452) were analyzed; the RSD values were in the range of 1.7–4.5% | Marine biota samples | — | [65] |
|
| |||||||||
| Total Hg | — | — | Direct mercury analyzer | 0.0027 μg/g | 0.002–0.15 μg | Recoveries of Hg were in the range of 98.9–106.1%; CRM (DORM-3) was analyzed |
Animal tissues | — | [66] |
|
| |||||||||
| Total Hg | — | — | Direct mercury analyzer (DMA-80) | 0–50 ng | SRM (NIST-1633b) and Rice fluor-NIES-10 (Japan) were analyzed | Human hair and nails | — | [57] | |
|
| |||||||||
| Total Hg | Platinum trap | — | Combustion mercury analyzer (MA 3000) | 0.00027 μg/g | — | CRMs (INCT-PVTL-6) and STRP-IS3 were analyzed | Tobacco samples | — | [67] |
|
| |||||||||
| Total Hg | Colorimetric | Lysine | Anisotropic gold nanoparticles | 27 pM | 0.01–1.0 nM | — | Deionized and tap waters | No interference of As3+, Cr3+, Cd2+, Pb2+, Ni2+, Zn2+, and Ba2+ was observed | [51] |
|
| |||||||||
| Total Hg | US EPA method 7473 | Direct mercury analyzer | — | — | Method was compared with TD-AAS results | Sediments | — | [93] | |
|
| |||||||||
| Hg(II) | Electrochemiluminescent | Graphene coupled quantum dots | MPI-A multifunctional electrochemical analytical system | 0.0003 μM | 0.2–5 μM | The RSD values in the determination of Hg(II) real samples were in the range of 2.4–7.5% | Spiked and real water samples | No interference of Cu2+, Pb2+, Ni2+, and Cd2+ was observed | [145] |
|
| |||||||||
| Hg(II) | Photoelectrochemical | CdS quantum dots | Atomic force microscope | 6.0 × 10−10 M | 3.0 × 10−9–1.0 × 10−7 M | — | — | 100-fold Cr3+, Fe3+, Pb2+, Cd2+, Cu2+, Mn2+, Zn2+, Al3+, and Co3+ did not interfere | [87] |
Regarding the usage of analytical techniques in the determination studies of mercury, as presented in Figure 1, 52.00% used spectrometric techniques, such as AAS, AFS, ICP-OES, ICP-MS, spectrometry, and spectrofluorometry, 30.00% used electrochemical techniques, such as a voltammetry and potentiometry, and 15.00% used miscellaneous techniques.
Figure 1.

Determination and speciation of mercury using various analytical techniques.
The analytical variables of merit, such as limits of detection and linearity, quality control and assurance studies, applicability to natural samples, and interference, are indicative of the validity of the method. Detection limit indicates the lowest level of analyte that can be detected using the method. A few studies [45–51] reported lowest levels down to picograms of mercury. Methods used in these studies can be considered highly sensitive because of their low detection limits. The lower detection limits were primarily obtained with ICP-MS and CV-AFS. On the other hand, spectrophotometers and spectrofluorometers can provide reasonable sensitivity, and they are inexpensive compared to ICP instruments. Linearity describes the range within which the method can determine analyte concentrations. Most of the electrochemical methods and spectrophotometry and spectrofluorometry methods determined the linearity range of analyte concentrations.
Two very important analytical parameters, in the determination of mercury, are quality assurance (QA) and quality control (QC). The validity and reliability of the data produced by the researchers depend on the quantification of these variables. Quality assurance studies can be performed by testing the accuracy of the data obtained against standard reference materials (SRMs) provided by the National Institute of Standards and Technology (NIST, USA) or certified reference materials (CRMs) provided by various reputable institutes or organizations. Quality control can be determined by measuring the precision of the data (repeatability and sensitivity) obtained by each method. The precision of the data can be obtained in many ways, such as the analysis of replicates, interlab comparison of data, and relative standard deviation (RSD) of blank or standard material analysis [52]. Regarding QA, in the reviewed papers, a few [45, 53–57] reported the analysis of SRMs to compare with the results of their methods. The results obtained with the measuring of SRMs give validity to the obtained data. The other alternative to measure the accuracy of the data is the analysis of CRMs. A large number of studies [45, 47, 55, 56, 58–67] reported the analysis of CRMs to validate their data. Regarding QC, most of the studies reported the RSD values for replicate sample analysis and/or standard materials analysis. Overall, most researchers were aware of the quality of their data, whereas a few [46, 51, 68–87] did not report any QA or QC values, which negatively affected the reliability of their data.
The validity of analytical methods can be enhanced by applicability to natural samples. Regarding the analysis of natural samples, most of the reviewed papers analyzed water samples, such as those from rivers, lakes, seas, groundwater, and spiked water and wastewater. Following water samples, the most frequently analyzed material for mercury was seafood samples, such as fish, shrimp, and seaweed. A few studies reported the determination of mercury in various environmental samples, such as petroleum hydrocarbons [88], human hair [89, 90], phosphate fertilizers [53], glycerin [91], sediments [55, 92, 93], cosmetics [94], living cells [66, 95], and tobacco [67]. However, a few methods [74, 81, 87, 96–101] did not report their applicability to natural samples. We found that very few authors [46] determined mercury content in the atmosphere. Because of the difficulty in sampling and analysis, most authors did not address this issue.
Another important aspect of analytical parameters of the methods is interference. Interference of other ions in the determination of mercury levels is very important, particularly when those methods are applied to the analysis of natural samples. Natural samples are typically a complex of different ions; the selectivity of the method is very important in the determination of mercury in environmental samples. In this respect, electrochemical methods had a higher degree of selectivity and did not suffer from interference from other ions. Determination of the level of interference was not performed by a large number of authors [45–47, 53, 55, 59, 60, 62, 63, 69, 88, 91, 92, 102–106] who determined mercury with spectrometric instruments. However, those using electrochemical instruments, spectrophotometers, or spectrofluorometers largely reported the level of interfering ion(s).
Toxicity of mercury depends upon its chemical form. For example, methylmercury is more toxic than inorganic mercury. Speciation studies revealed the exact toxicity of mercury in environmental samples. However, very few authors [88, 90, 92, 107] reported the speciation of mercury, and most authors determined the level of inorganic mercury. More than 90% of studies using electrochemical methods or spectrophotometry and spectrofluorometry techniques determined divalent inorganic mercury and did not report speciation. However, a few reported [51, 53, 55, 57, 58, 60–62, 66–68, 85, 86, 93, 102–106, 108, 109] total mercury content in various samples, which does not accurately predict toxicity based on its concentration.
4. Conclusions
The present study reviewed research articles published in recent years (2015-2016) involving determination and speciation of mercury using various analytical instruments. Approximately 100 research papers were reviewed and all the analytical parameters established in their studies were tabulated. Our study concluded that most of researchers used spectrometric instruments for the determination of mercury in different environmental samples. We addressed the quality of the data based on reported QA and QC data by the authors. Another important finding from this review was that most researchers measured inorganic mercury or total mercury, whereas only a few reported speciation of mercury. Speciation studies are very important in the accurate prediction of the toxicity of the mercury in the environment because mercury toxicity depends on its chemical form. Because of the difficulty in sampling and analysis, most researchers did not report the concentrations of mercury in the atmosphere. We conclude by stating that speciation studies and the determination of mercury in the atmosphere should receive greater attention in the future.
Conflicts of Interest
The authors declare that there are no conflicts of interest regarding the publication of this article.
References
- 1.Arantes F. P., Savassi L. A., Santos H. B., Gomes M. V. T., Bazzoli N. Bioaccumulation of mercury, cadmium, zinc, chromium, and lead in muscle, liver, and spleen tissues of a large commercially valuable catfish species from Brazil. Anais da Academia Brasileira de Ciências. 2016;88(1):137–147. doi: 10.1590/0001-3765201620140434. [DOI] [PubMed] [Google Scholar]
- 2.Bastos W. R., Dórea J. G., Bernardi J. V. E., et al. Sex-related mercury bioaccumulation in fish from the Madeira River, Amazon. Environmental Research. 2016;144:73–80. doi: 10.1016/j.envres.2015.11.001. [DOI] [PubMed] [Google Scholar]
- 3.Øverjordet I. B., Gabrielsen G. W., Berg T., et al. Effect of diet, location and sampling year on bioaccumulation of mercury, selenium and cadmium in pelagic feeding seabirds in Svalbard. Chemosphere. 2015;122:14–22. doi: 10.1016/j.chemosphere.2014.10.060. [DOI] [PubMed] [Google Scholar]
- 4.Li J., Zhou Q., Yuan G., He X., Xie P. Mercury bioaccumulation in the food web of Three Gorges Reservoir (China): tempo-spatial patterns and effect of reservoir management. Science of the Total Environment. 2015;527-528:203–210. doi: 10.1016/j.scitotenv.2015.04.115. [DOI] [PubMed] [Google Scholar]
- 5.Dang F., Zhao J., Greenfield B. K., et al. Soil geochemistry and digestive solubilization control mercury bioaccumulation in the earthworm Pheretima guillemi. Journal of Hazardous Materials. 2015;292:44–51. doi: 10.1016/j.jhazmat.2015.03.012. [DOI] [PubMed] [Google Scholar]
- 6.Gosnell K. J., Mason R. P. Mercury and methylmercury incidence and bioaccumulation in plankton from the central Pacific Ocean. Marine Chemistry. 2015;177:772–780. doi: 10.1016/j.marchem.2015.07.005. [DOI] [Google Scholar]
- 7.Banaee M., Mohammadipour S., Madhani S. Effects of sublethal concentrations of permethrin on bioaccumulation of cadmium in zebra cichlid (Cichlasoma nigrofasciatum) Toxicological & Environmental Chemistry. 2015;97(2):200–207. doi: 10.1080/02772248.2015.1031668. [DOI] [Google Scholar]
- 8.Ouédraogo O., Chételat J., Amyot M. Bioaccumulation and trophic transfer of mercury and selenium in African sub-tropical fluvial reservoirs food webs (Burkina Faso) PLoS ONE. 2015;10(4) doi: 10.1371/journal.pone.0123048.e0123048 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Donald D. B., Wissel B., Anas M. U. M. Species-specific mercury bioaccumulation in a diverse fish community. Environmental Toxicology and Chemistry. 2015;34(12):2846–2855. doi: 10.1002/etc.3130. [DOI] [PubMed] [Google Scholar]
- 10. https://people.uwec.edu/piercech/Hg/mercury_water/ursoces.htm; accessed on 8th August, 2016.
- 11.Zhuang W., Gao X. Distributions, sources and ecological risk assessment of arsenic and mercury in the surface sediments of the southwestern coastal Laizhou Bay, Bohai Sea. Marine Pollution Bulletin. 2015;99(1-2):320–327. doi: 10.1016/j.marpolbul.2015.07.037. [DOI] [PubMed] [Google Scholar]
- 12.Kwon S. Y., Blum J. D., Nadelhoffer K. J., Timothy Dvonch J., Tsui M. T.-K. Isotopic study of mercury sources and transfer between a freshwater lake and adjacent forest food web. Science of the Total Environment. 2015;532:220–229. doi: 10.1016/j.scitotenv.2015.06.012. [DOI] [PubMed] [Google Scholar]
- 13.Xu J., Bravo A. G., Lagerkvist A., Bertilsson S., Sjöblom R., Kumpiene J. Sources and remediation techniques for mercury contaminated soil. Environment International. 2014;74:42–53. doi: 10.1016/j.envint.2014.09.007. [DOI] [PubMed] [Google Scholar]
- 14.Shamsipur M., Rajabi H. R., Beyzavi M. H., Sharghi H. Bulk polymer nanoparticles containing a tetrakis(3-hydroxyphenyl)porphyrin for fast and highly selective separation of mercury ions. Microchimica Acta. 2013;180(9-10):791–799. doi: 10.1007/s00604-013-0983-x. [DOI] [Google Scholar]
- 15.Rajabi H. R., Roushani M., Shamsipur M. Development of a highly selective voltammetric sensor for nanomolar detection of mercury ions using glassy carbon electrode modified with a novel ion imprinted polymeric nanobeads and multi-wall carbon nanotubes. Journal of Electroanalytical Chemistry. 2013;693:16–22. doi: 10.1016/j.jelechem.2013.01.003. [DOI] [Google Scholar]
- 16.Han J.-S., Seo Y.-S., Kim M.-K., Holsen T. M., Yi S.-M. Total atmospheric mercury deposition in forested areas in South Korea. Atmospheric Chemistry and Physics. 2016;16(12):7653–7662. doi: 10.5194/acp-16-7653-2016. [DOI] [Google Scholar]
- 17.Xu X., Liu N., Landis M. S., Feng X., Qiu G. Characteristics and distributions of atmospheric mercury emitted from anthropogenic sources in Guiyang, southwestern China. Acta Geochimica. 2016;35(3):240–250. doi: 10.1007/s11631-016-0111-9. [DOI] [Google Scholar]
- 18.Yoshida M., Lee J., Shimizu-Furusawa H., Satoh M., Watanabe C. Neurobehavioral toxicity related to the exposure of weaning mice to low-level mercury vapor and methylmercury and influence of aging. Fundamental Toxicological Sciences. 2016;3(4):185–193. doi: 10.2131/fts.3.185. [DOI] [Google Scholar]
- 19.Bucio L., Souza V., Albores A., et al. Cadmium and mercury toxicity in a human fetal hepatic cell line (WRL-68 cells) Toxicology. 1995;102(3):285–299. doi: 10.1016/0300-483x(95)03095-w. [DOI] [PubMed] [Google Scholar]
- 20.Farina M., Avila D. S., Da Rocha J. B. T., Aschner M. Metals, oxidative stress and neurodegeneration: A focus on iron, manganese and mercury. Neurochemistry International. 2013;62(5):575–594. doi: 10.1016/j.neuint.2012.12.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Al-Batanony M. A., Abdel-Rasul G. M., Abu-Salem M. A., Al-Dalatony M. M., Allam H. K. Occupational exposure to mercury among workers in a fluorescent lamp factory, Quisna industrial zone, Egypt. International Journal of Occupational Medicine and Environmental Health. 2013;4(3):149–156. [PubMed] [Google Scholar]
- 22.Carvalho C. M. L., Chew E.-H., Hashemy S. I., Lu J., Holmgren A. Inhibition of the human thioredoxin system: A molecular mechanism of mercury toxicity. The Journal of Biological Chemistry. 2008;283(18):11913–11923. doi: 10.1074/jbc.M710133200. [DOI] [PubMed] [Google Scholar]
- 23.Bernhoft R. A. Mercury toxicity and treatment: a review of the literature. Journal of Environmental and Public Health. 2012;2012:10. doi: 10.1155/2012/460508.460508 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Tonazzi A., Giangregorio N., Console L., et al. Mitochondrial Carnitine/Acylcarnitine Transporter, a Novel Target of Mercury Toxicity. Chemical Research in Toxicology. 2015;28(5):1015–1022. doi: 10.1021/acs.chemrestox.5b00050. [DOI] [PubMed] [Google Scholar]
- 25.Xu F., Farkas S., Kortbeek S., et al. Mercury-induced toxicity of rat cortical neurons is mediated through N-methyl-D-Aspartate receptors. Molecular Brain. 2012;5, article 30 doi: 10.1186/1756-6606-5-30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Dórea J. G., Farina M., Rocha J. B. T. Toxicity of ethylmercury (and Thimerosal): a comparison with methylmercury. Journal of Applied Toxicology. 2013;33(8):700–711. doi: 10.1002/jat.2855. [DOI] [PubMed] [Google Scholar]
- 27.Houston M. C. Role of mercury toxicity in hypertension, cardiovascular disease, and stroke. The Journal of Clinical Hypertension. 2011;13(8):621–627. doi: 10.1111/j.1751-7176.2011.00489.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Rice K. M., Walker E. M., Jr., Wu M., Gillette C., Blough E. R. Environmental mercury and its toxic effects. Journal of Preventive Medicine & Public Health. 2014;47(2):74–83. doi: 10.3961/jpmph.2014.47.2.74. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Patra M., Sharma A. Mercury toxicity in plants. The Botanical Review. 2000;66(3):379–422. doi: 10.1007/BF02868923. [DOI] [Google Scholar]
- 30.Cargnelutti D., Tabaldi L. A., Spanevello R. M., et al. Mercury toxicity induces oxidative stress in growing cucumber seedlings. Chemosphere. 2006;65(6):999–1006. doi: 10.1016/j.chemosphere.2006.03.037. [DOI] [PubMed] [Google Scholar]
- 31.Chen J., Yang Z. M. Mercury toxicity, molecular response and tolerance in higher plants. BioMetals. 2012;25(5):847–857. doi: 10.1007/s10534-012-9560-8. [DOI] [PubMed] [Google Scholar]
- 32.Hanna D. E. L., Solomon C. T., Poste A. E., Buck D. G., Chapman L. J. A review of mercury concentrations in freshwater fishes of Africa: Patterns and predictors. Environmental Toxicology and Chemistry. 2015;34(2):215–223. doi: 10.1002/etc.2818. [DOI] [PubMed] [Google Scholar]
- 33.Ferreira S. L. C., Lemos V. A., Silva L. O. B., et al. Analytical strategies of sample preparation for the determination of mercury in food matrices—a review. Microchemical Journal. 2015;121:227–236. doi: 10.1016/j.microc.2015.02.012. [DOI] [Google Scholar]
- 34.Chansuvarn W., Tuntulani T., Imyim A. Colorimetric detection of mercury(II) based on gold nanoparticles, fluorescent gold nanoclusters and other gold-based nanomaterials. TrAC - Trends in Analytical Chemistry. 2015;65:83–96. doi: 10.1016/j.trac.2014.10.013. [DOI] [Google Scholar]
- 35.Ariya P. A., Amyot M., Dastoor A., et al. Mercury physicochemical and biogeochemical transformation in the atmosphere and at atmospheric interfaces: a review and future directions. Chemical Reviews. 2015;115(10):3760–3802. doi: 10.1021/cr500667e. [DOI] [PubMed] [Google Scholar]
- 36.Shrivastava S., Shrivastav A., Sharma J. Detoxification mechanisms of mercury toxicity in plants: a review. Recent Advances in Biology and Medicine. 2015;01:p. 60. doi: 10.18639/RABM.2015.01.196308. [DOI] [Google Scholar]
- 37.Duarte K., Justino C. I. L., Freitas A. C., Gomes A. M. P., Duarte A. C., Rocha-Santos T. A. P. Disposable sensors for environmental monitoring of lead, cadmium and mercury. TrAC Trends in Analytical Chemistry. 2015;64:183–190. doi: 10.1016/j.trac.2014.07.006. [DOI] [Google Scholar]
- 38.Jagtap R., Maher W. Measurement of mercury species in sediments and soils by HPLC-ICPMS. Microchemical Journal. 2015;121:65–98. doi: 10.1016/j.microc.2015.01.010. [DOI] [Google Scholar]
- 39.Gustin M. S., Amos H. M., Huang J., Miller M. B., Heidecorn K. Measuring and modeling mercury in the atmosphere: a critical review. Atmospheric Chemistry and Physics. 2015;15(10):5697–5713. doi: 10.5194/acp-15-5697-2015. [DOI] [Google Scholar]
- 40.McLagan D. S., Mazur M. E., Mitchell C. P., Wania F. Passive air sampling of gaseous elemental mercury: a critical review. Atmospheric Chemistry and Physics Discussions. 2015;15(23):34605–34645. doi: 10.5194/acpd-15-34605-2015. [DOI] [Google Scholar]
- 41.Jackson B. P., Punshon T. Recent advances in the measurement of arsenic, cadmium, and mercury in rice and other foods. Current Environmental Health Reports. 2015;2(1):15–24. doi: 10.1007/s40572-014-0035-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Duan J., Zhan J. Recent developments on nanomaterials-based optical sensors for Hg2+ detection. Science China Materials. 2015;58(3):223–240. doi: 10.1007/s40843-015-0031-8. [DOI] [Google Scholar]
- 43.Sun Z., Du J., Jing C. Recent progress in detection of mercury using surface enhanced Raman spectroscopy - a review. Journal of Environmental Sciences. 2016;39:134–143. doi: 10.1016/j.jes.2015.11.009. [DOI] [PubMed] [Google Scholar]
- 44.Suvarapu L. N., Baek S.-O. Recent developments in the speciation and determination of mercury using various analytical techniques. Journal of Analytical Methods in Chemistry. 2015;2015:18. doi: 10.1155/2015/372459.372459 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Chen X., Balasubramanian R., Zhu Q., et al. Characteristics of atmospheric particulate mercury in size-fractionated particles during haze days in Shanghai. Atmospheric Environment. 2016;131:400–408. doi: 10.1016/j.atmosenv.2016.02.019. [DOI] [Google Scholar]
- 46.Marumoto K., Imai S. Determination of dissolved gaseous mercury in seawater of Minamata Bay and estimation for mercury exchange across air-sea interface. Marine Chemistry. 2015;168:9–17. doi: 10.1016/j.marchem.2014.09.007. [DOI] [Google Scholar]
- 47.Brombach C.-C., Chen B., Corns W. T., Feldmann J., Krupp E. M. Methylmercury in water samples at the pg/L level by online preconcentration liquid chromatography cold vapor-atomic fluorescence spectrometry. Spectrochimica Acta Part B: Atomic Spectroscopy. 2015;105:103–108. doi: 10.1016/j.sab.2014.09.014. [DOI] [Google Scholar]
- 48.Zhang Y., Zeng G. M., Tang L., et al. Electrochemical sensor based on electrodeposited graphene-au modified electrode and nanoau carrier amplified signal strategy for attomolar mercury detection. Analytical Chemistry. 2015;87(2):989–996. doi: 10.1021/ac503472p. [DOI] [PubMed] [Google Scholar]
- 49.Liu X., Wu Z., Zhang Q., Zhao W., Zong C., Gai H. Single gold nanoparticle-based colorimetric detection of picomolar mercury ion with dark-field microscopy. Analytical Chemistry. 2016;88(4):2119–2124. doi: 10.1021/acs.analchem.5b03653. [DOI] [PubMed] [Google Scholar]
- 50.Huang R.-F., Liu H.-X., Gai Q.-Q., Liu G.-J., Wei Z. A facile and sensitive electrochemiluminescence biosensor for Hg2+ analysis based on a dual-function oligonucleotide probe. Biosensors and Bioelectronics. 2015;71:194–199. doi: 10.1016/j.bios.2015.04.038. [DOI] [PubMed] [Google Scholar]
- 51.Chaudhary A., Dwivedi C., Chawla M., Gupta A., Nandi C. K. Lysine and dithiothreitol promoted ultrasensitive optical and colorimetric detection of mercury using anisotropic gold nanoparticles. Journal of Materials Chemistry C. 2015;3(27):6962–6965. doi: 10.1039/c5tc01397f. [DOI] [Google Scholar]
- 52.Miller J. C., Miller J. N. Statistics for Analytical Chemistry. 2nd. England: Ellis Horwood Limited; 1988. [Google Scholar]
- 53.De Oliveira Souza S., François L. L., Borges A. R., Vale M. G. R., Araujo R. G. O. Determination of copper and mercury in phosphate fertilizers employing direct solid sampling analysis and high resolution continuum source graphite furnace atomic absorption spectrometry. Spectrochimica Acta Part B: Atomic Spectroscopy. 2015;114:58–64. doi: 10.1016/j.sab.2015.10.003. [DOI] [Google Scholar]
- 54.Ghorbani-Kalhor E., Hosseinzadeh-Khanmiri R., Abolhasani J., Babazadeh M., Hassanpour A. Determination of mercury(II) ions in seafood samples after extraction and preconcentration by a novel functionalized magnetic metal-organic framework nanocomposite. Journal of Separation Science. 2015;38(7):1179–1186. doi: 10.1002/jssc.201401320. [DOI] [PubMed] [Google Scholar]
- 55.Mandjukov P., Orani A. M., Han E., Vassileva E. Determination of total mercury for marine environmental monitoring studies by solid sampling continuum source high resolution atomic absorption spectrometry. Spectrochimica Acta Part B: Atomic Spectroscopy. 2015;103-104:24–33. doi: 10.1016/j.sab.2014.11.006. [DOI] [Google Scholar]
- 56.Soares L. C., Egreja Filho F. B., Linhares L. A., Windmoller C. C., Yoshida M. I. Accumulation and oxidation of elemental mercury in tropical soils. Chemosphere. 2015;134:181–191. doi: 10.1016/j.chemosphere.2015.04.020. [DOI] [PubMed] [Google Scholar]
- 57.Vicent A., Cervera M. L., Rubio A. M. Fast and accurate methodology for direct mercury determination in hair and nails. Sample Preparation. 2015;2(1):91–98. doi: 10.1515/sampre-2015-0006. [DOI] [Google Scholar]
- 58.Li Y., Zhu Z., Zheng H., Jin L., Hu S. Significant signal enhancement of dielectric barrier discharge plasma induced vapor generation by using non-ionic surfactants for determination of mercury and cadmium by atomic fluorescence spectrometry. Journal of Analytical Atomic Spectrometry. 2016;31(2):383–389. doi: 10.1039/c5ja00300h. [DOI] [Google Scholar]
- 59.Pietilä H., Perämäki P., Piispanen J., et al. Determination of low methylmercury concentrations in peat soil samples by isotope dilution GC-ICP-MS using distillation and solvent extraction methods. Chemosphere. 2015;124(1):47–53. doi: 10.1016/j.chemosphere.2014.11.001. [DOI] [PubMed] [Google Scholar]
- 60.Panichev N. A., Panicheva S. E. Determination of total mercury in fish and sea products by direct thermal decomposition atomic absorption spectrometry. Food Chemistry. 2015;166:432–441. doi: 10.1016/j.foodchem.2014.06.032. [DOI] [PubMed] [Google Scholar]
- 61.Sánchez R., Snell J., Held A., Emons H. Development and validation of a method for mercury determination in seawater for the process control of a candidate certified reference material. Analytical and Bioanalytical Chemistry. 2015;407(21):6569–6574. doi: 10.1007/s00216-015-8833-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Fernández-Martínez R., Rucandio I., Gómez-Pinilla I., Borlaf F., García F., Larrea M. T. Evaluation of different digestion systems for determination of trace mercury in seaweeds by cold vapour atomic fluorescence spectrometry. Journal of Food Composition and Analysis. 2015;38:7–12. doi: 10.1016/j.jfca.2014.10.003. [DOI] [Google Scholar]
- 63.Döker S., Boşgelmez I. I. Rapid extraction and reverse phase-liquid chromatographic separation of mercury(II) and methylmercury in fish samples with inductively coupled plasma mass spectrometric detection applying oxygen addition into plasma. Food Chemistry. 2015;184:147–153. doi: 10.1016/j.foodchem.2015.03.067.17314 [DOI] [PubMed] [Google Scholar]
- 64.Chen Z., Zhang C., Ma H., et al. A non-aggregation spectrometric determination for mercury ions based on gold nanoparticles and thiocyanuric acid. Talanta. 2015;134:603–606. doi: 10.1016/j.talanta.2014.11.065. [DOI] [PubMed] [Google Scholar]
- 65.Azemard S., Vassileva E. Determination of methylmercury in marine biota samples with advanced mercury analyzer: method validation. Food Chemistry. 2015;176:367–375. doi: 10.1016/j.foodchem.2014.12.085. [DOI] [PubMed] [Google Scholar]
- 66.Ribeiro R. F. L., Germano A. Development and validation of a method for the determination of Hg in animal tissues (equine muscle, bovine kidney and swine kidney, and poultry muscle) by direct mercury analysis (DMA) Microchemical Journal. 2015;121:237–243. doi: 10.1016/j.microc.2015.03.005. [DOI] [Google Scholar]
- 67.Fresquez M. R., Gonzalez-Jimenez N., Gray N., Watson C. H., Pappas R. S. High-throughput determination of mercury in tobacco and mainstream smoke from little cigars. Journal of Analytical Toxicology. 2015;39(7):545–550. doi: 10.1093/jat/bkv069. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Mortazavi A., Hatamikia M., Bahmani M., Hassanzadazar H. Heavy metals (mercury, lead and cadmium) determination of 17 species of fish marketed in Khorramabad city, west of Iran. Journal of Chemical Health Risks. 2016;6(1):41–48. [Google Scholar]
- 69.Minh N. P. Determination of mercury in seafood by atomic emission spectrophotometry. Bulletin of Environment, Pharmacology and Life Sciences. 2015;4(7):41–47. [Google Scholar]
- 70.Bhatt K. D., Vyas D. J., Makwana B. A., Darjee S. M., Jain V. K., Shah H. Turn-on fluorescence probe for selective detection of Hg(II) by calixpyrrole hydrazide reduced silver nanoparticle: Application to real water sample. Chinese Chemical Letters. 2016;27(5):731–737. doi: 10.1016/j.cclet.2016.01.012. [DOI] [Google Scholar]
- 71.Geng Z., Zhang H., Xiong Q., Zhang Y., Zhao H., Wang G. A fluorescent chitosan hydrogel detection platform for the sensitive and selective determination of trace mercury(II) in water. Journal of Materials Chemistry A. 2015;3(38):19455–19460. doi: 10.1039/c5ta05610a. [DOI] [Google Scholar]
- 72.Velmurugan K., Nandhakumar R. Binol based 'turn on' fluorescent chemosensor for mercury ion. Journal of Luminescence. 2015;162:8–13. doi: 10.1016/j.jlumin.2015.01.039. [DOI] [Google Scholar]
- 73.Erdemir S., Kocyigit O., Malkondu S. Detection of Hg2+ ion in aqueous media by new fluorometric and colorimetric sensor based on triazole-rhodamine. Journal of Photochemistry and Photobiology A: Chemistry. 2015;309, article no. 9880:15–21. doi: 10.1016/j.jphotochem.2015.04.017. [DOI] [Google Scholar]
- 74.Hu J., Li J., Qi J., Chen J. Highly selective and effective mercury(II) fluorescent sensors. New Journal of Chemistry. 2015;39(2):843–848. doi: 10.1039/c4nj01147c. [DOI] [Google Scholar]
- 75.Shi D., Yan F., Wang M., et al. Rhodamine derivative functionalized chitosan as efficient sensor and adsorbent for mercury(II) detection and removal. Materials Research Bulletin. 2015;70:958–964. doi: 10.1016/j.materresbull.2015.06.046. [DOI] [Google Scholar]
- 76.Guo Y., An J., Tang H., Peng M., Suzenet F. Selective and 'turn-off' fluorimetric detection of mercury(II) based on coumarinyldithiolane and coumarinyldithiane in aqueous solution. Materials Research Bulletin. 2015;63:155–163. doi: 10.1016/j.materresbull.2014.12.015. [DOI] [Google Scholar]
- 77.Wang M., Meng G., Huang Q. Spinach-extracted chlorophyll-a modified peanut shell as fluorescence sensors for selective detection of Hg2+ in water. Sensors and Actuators B: Chemical. 2015;209:237–241. doi: 10.1016/j.snb.2014.11.117. [DOI] [Google Scholar]
- 78.Guan J., Wang Y.-C., Gunasekaran S. Using L-Arginine-Functionalized Gold Nanorods for Visible Detection of Mercury(II) Ions. Journal of Food Science. 2015;80(4):N828–N833. doi: 10.1111/1750-3841.12811. [DOI] [PubMed] [Google Scholar]
- 79.Guan Q., Xiong W., Zhou L., Liu S. Facile synthesis of nitrogen-doped porous carbon-gold hybrid nanocomposite for mercury(II) ion electrochemical determination. Electroanalysis. 2016;28(1):133–139. doi: 10.1002/elan.201500481. [DOI] [Google Scholar]
- 80.Wang N., Lin M., Dai H., Ma H. Functionalized gold nanoparticles/reduced graphene oxide nanocomposites for ultrasensitive electrochemical sensing of mercury ions based on thymine-mercury-thymine structure. Biosensors and Bioelectronics. 2016;79:320–326. doi: 10.1016/j.bios.2015.12.056. [DOI] [PubMed] [Google Scholar]
- 81.Jiang B., Yu L., Li F., Xie J. A dual functional electrochemical 'on-off' switch sensor for the detection of mercury(II) and melamine. Sensors and Actuators B: Chemical. 2015;212:446–450. doi: 10.1016/j.snb.2015.02.034. [DOI] [Google Scholar]
- 82.Bahrami A., Besharati-Seidani A., Abbaspour A., Shamsipur M. A highly selective voltammetric sensor for nanomolar detection of mercury ions using a carbon ionic liquid paste electrode impregnated with novel ion imprinted polymeric nanobeads. Materials Science and Engineering C: Materials for Biological Applications. 2015;48:205–212. doi: 10.1016/j.msec.2014.12.005. [DOI] [PubMed] [Google Scholar]
- 83.Ratner N., Mandler D. Electrochemical detection of low concentrations of mercury in water using gold nanoparticles. Analytical Chemistry. 2015;87(10):5148–5155. doi: 10.1021/ac504584f. [DOI] [PubMed] [Google Scholar]
- 84.Xie Y.-L., Zhao S.-Q., Ye H.-L., Yuan J., Song P., Hu S.-Q. Graphene/CeO2 hybrid materials for the simultaneous electrochemical detection of cadmium(II), lead(II), copper(II), and mercury(II) Journal of Electroanalytical Chemistry. 2015;757:235–242. doi: 10.1016/j.jelechem.2015.09.043. [DOI] [Google Scholar]
- 85.Bhanjana G., Dilbaghi N., Kumar R., Kumar S. Zinc oxide quantum dots as efficient electron mediator for ultrasensitive and selective electrochemical sensing of mercury. Electrochimica Acta. 2015;178:361–367. doi: 10.1016/j.electacta.2015.07.113.25395 [DOI] [Google Scholar]
- 86.Fard N. J. H., Ravanbakhsh M., Ramezani Z., Ahmadi M., Angali K. A., Javid A. Z. Determination of mercury and vanadium concentration in Johnius belangerii (C) fish in Musa estuary in Persian Gulf. Marine Pollution Bulletin. 2015;97(1-2):499–505. doi: 10.1016/j.marpolbul.2015.05.036. [DOI] [PubMed] [Google Scholar]
- 87.Han D.-M., Jiang L.-Y., Tang W.-Y., Xu J.-J., Chen H.-Y. Photoelectrochemical determination of inorganic mercury ions based on energy transfer between CdS quantum dots and Au nanoparticles. Electrochemistry Communications. 2015;51:72–75. doi: 10.1016/j.elecom.2014.12.002. [DOI] [Google Scholar]
- 88.Gajdosechova Z., Boskamp M. S., Lopez-Linares F., Feldmann J., Krupp E. M. Hg speciation in petroleum hydrocarbons with emphasis on the reactivity of Hg particles. Energy & Fuels. 2016;30(1):130–137. doi: 10.1021/acs.energyfuels.5b02080. [DOI] [Google Scholar]
- 89.Wen X., Gao Y., Wu P., Tan Z., Zheng C., Hou X. Chemical vapor generation from an ionic liquid using a solid reductant: determination of Hg, As and Sb by atomic fluorescence spectrometry. Journal of Analytical Atomic Spectrometry. 2016;31(2):415–422. doi: 10.1039/c5ja00323g. [DOI] [Google Scholar]
- 90.Li P., He M., Chen B., Hu B. Automated dynamic hollow fiber liquid-liquid-liquid microextraction combined with capillary electrophoresis for speciation of mercury in biological and environmental samples. Journal of Chromatography A. 2015;1415:48–56. doi: 10.1016/j.chroma.2015.08.062. [DOI] [PubMed] [Google Scholar]
- 91.Lisboa M. T., Clasen C. D., Oreste E. Q., Schwingel Ribeiro A., Vieira M. A. Comparison between vapor generation methods coupled to atomic absorption spectrometry for determination of Hg in glycerin samples. ENERGY & FUELS. 2015;29(3):1635–1640. doi: 10.1021/ef501399d. [DOI] [Google Scholar]
- 92.Leng G., Chen W., Wang Y. Speciation analysis of mercury in sediments using ionic-liquid-based vortex-assisted liquid-liquid microextraction combined with high-performance liquid chromatography and cold vapor atomic fluorescence spectrometry. Journal of Separation Science. 2015;38(15):2684–2691. doi: 10.1002/jssc.201500083. [DOI] [PubMed] [Google Scholar]
- 93.Chen J., Chakravarty P., Davidson G. R., et al. Simultaneous determination of mercury and organic carbon in sediment and soils using a direct mercury analyzer based on thermal decomposition-atomic absorption spectrophotometry. Analytica Chimica Acta. 2015;871:9–17. doi: 10.1016/j.aca.2015.03.011. [DOI] [PubMed] [Google Scholar]
- 94.Prasertboonyai K., Liawraungrath B., Pojanakaroon T., Liawraungrath S. Mercury(II) determination in commercial cosmetics and local Thai traditional medicines by flow injection spectrophotometry. International Journal of Cosmetic Science. 2016;38(1):68–76. doi: 10.1111/ics.12259. [DOI] [PubMed] [Google Scholar]
- 95.Shellaiah M., Rajan Y. C., Balu P., Murugan A. A pyrene based Schiff base probe for selective fluorescence turn-on detection of Hg2+ ions with live cell application. New Journal of Chemistry. 2015;39(4):2523–2531. doi: 10.1039/c4nj02367f. [DOI] [Google Scholar]
- 96.He X., Zhu S., Chen H., Wang Y., Li H. A new fluorescent turn-on chemodosimeter for mercury(II) based on dithioacetal-substituted triphenylimidazole. Journal of Luminescence. 2016;173:218–222. doi: 10.1016/j.jlumin.2016.01.025. [DOI] [Google Scholar]
- 97.He X., Yang D., Chen H., Zheng W., Li H. A highly sensitive and reversible chemosensor for Hg2+ detection based on porphyrin-thymine conjugates. Journal of Molecular Recognition. 2015 doi: 10.1002/jmr.2442. [DOI] [PubMed] [Google Scholar]
- 98.Hien N. K., Bao N. C., Ai Nhung N. T., et al. A highly sensitive fluorescent chemosensor for simultaneous determination of Ag(I), Hg(II), and Cu(II) ions: design, synthesis, characterization and application. Dyes and Pigments. 2015;116:89–96. doi: 10.1016/j.dyepig.2015.01.014. [DOI] [Google Scholar]
- 99.Xing H., Xu J., Zhu X., et al. Highly sensitive simultaneous determination of cadmium (II), lead (II), copper (II), and mercury (II) ions on N-doped graphene modified electrode. Journal of Electroanalytical Chemistry. 2016;760:52–58. doi: 10.1016/j.jelechem.2015.11.043. [DOI] [Google Scholar]
- 100.Jovanovski V., Hrastnik N. I., Hočevar S. B. Copper film electrode for anodic stripping voltammetric determination of trace mercury and lead. Electrochemistry Communications. 2015;57:1–4. doi: 10.1016/j.elecom.2015.04.018. [DOI] [Google Scholar]
- 101.Lin H., Lou X., Zhong W., He S. Continuous monitoring of elemental mercury employing low-cost multimode diode lasers. Measurement Science and Technology. 2015;26(8) doi: 10.1088/0957-0233/26/8/085501.085501 [DOI] [Google Scholar]
- 102.Ali J., Kazi T. G., Afridi H. I., et al. The evaluation of sequentially extracted mercury fractions in Thar coal samples by using different extraction schemes. International Journal of Coal Geology. 2016;156:50–58. doi: 10.1016/j.coal.2016.02.003. [DOI] [Google Scholar]
- 103.Barbosa U. A., Silva L. O. B., Dos Santos I. F., Ferreira S. L. C., Dos Santos A. M. P. Determination of mercury in iron supplement using slurry sampling and cold vapor atomic absorption spectrometry. Current Analytical Chemistry. 2015;11(1):44–49. doi: 10.2174/1573411010666141031003321. [DOI] [Google Scholar]
- 104.De O Vicentino P., Brum D. M., Cassella R. J. Development of a method for total Hg determination in oil samples by cold vapor atomic absorption spectrometry after its extraction induced by emulsion breaking. Talanta. 2015;132:733–738. doi: 10.1016/j.talanta.2014.10.019. [DOI] [PubMed] [Google Scholar]
- 105.Torres D. P., Olivares I. R. B., Queiroz H. M. Estimate of the uncertainty in measurement for the determination of mercury in seafood by TDA AAS. Journal of Environmental Science and Health, Part B: Pesticides, Food Contaminants, and Agricultural Wastes. 2015;50(8):622–631. doi: 10.1080/03601234.2015.1028861. [DOI] [PubMed] [Google Scholar]
- 106.Torres D. P., Martins-Teixeira M. B., Cadore S., Queiroz H. M. Method validation for control determination of mercury in fresh fish and shrimp samples by solid sampling thermal decomposition/amalgamation atomic absorption spectrometry. Journal of Environmental Science and Health, Part B: Pesticides, Food Contaminants, and Agricultural Wastes. 2015;50(7):514–522. doi: 10.1080/03601234.2015.1018764. [DOI] [PubMed] [Google Scholar]
- 107.Samaddar P., Sen K. Discrimination for inorganic and organic mercury species by cloud point extraction of polyethylene glycol. Journal of Environmental Chemical Engineering (JECE) 2016;4(2):1862–1868. doi: 10.1016/j.jece.2016.03.010. [DOI] [Google Scholar]
- 108.Silva M. M., Jr., Bastos Silva L. O., Leao D. J., Lopes dos Santos W. N., Welz B., Costa Ferreira S. L. Determination of mercury in alcohol vinegar samples from Salvador, Bahia, Brazil. Food Control. 2015;47:623–627. doi: 10.1016/j.foodcont.2014.07.043. [DOI] [Google Scholar]
- 109.Rohman A., Wijayanti E. Development and validation of atomic absorption spectrometry for the determination of zinc and mercury analyzer for determination of mercury in cream cosmetics. Journal of Food and Pharmaceutical Sciences. 2015;3:23–26. [Google Scholar]
- 110.Lee J. J., Kim Y. S., Nam E., Lee S. Y., Lim M. H., Kim C. A pet-based fluorometric chemosensor for the determination of mercury(II) and pH, and hydrolysis reaction-based colorimetric detection of hydrogen sulfide. Dalton Transactions. 2016;45(13):5700–5712. doi: 10.1039/c6dt00147e. [DOI] [PubMed] [Google Scholar]
- 111.Francisco B. B. A., Rocha A. A., Grinberg P., Sturgeon R. E., Cassella R. J. Determination of inorganic mercury in petroleum production water by inductively coupled plasma optical emission spectrometry following photochemical vapor generation. Journal of Analytical Atomic Spectrometry. 2016;31(3):751–758. doi: 10.1039/c5ja00444f. [DOI] [Google Scholar]
- 112.Galimberti C., Corti I., Cressoni M., et al. Evaluation of mercury, cadmium and lead levels in fish and fishery products imported by air in North Italy from extra-European Union Countries. Food Control. 2015;60:329–337. doi: 10.1016/j.foodcont.2015.08.009. [DOI] [Google Scholar]
- 113.Pang Y., Rong Z., Xiao R., Wang S. 'Turn on' and label-free core - Shell Ag@SiO2 nanoparticles-based metal-enhanced fluorescent (MEF) aptasensor for Hg2+ Scientific Reports. 2015;5, article 9451 doi: 10.1038/srep09451. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114.Zargoosh K., Babadi F. F. Highly selective and sensitive optical sensor for determination of Pb2+and Hg2+ions based on the covalent immobilization of dithizone on agarose membrane. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy. 2015;137:105–110. doi: 10.1016/j.saa.2014.08.043. [DOI] [PubMed] [Google Scholar]
- 115.Rajabi H. R., Shamsipur M., Zahedi M. M., Roushani M. On-line flow injection solid phase extraction using imprinted polymeric nanobeads for the preconcentration and determination of mercury ions. Chemical Engineering Journal. 2015;259:330–337. doi: 10.1016/j.cej.2014.08.025. [DOI] [Google Scholar]
- 116.Ding X., Qu L., Yang R., Zhou Y., Li J. A highly selective and simple fluorescent sensor for mercury (II) ion detection based on cysteamine-capped CdTe quantum dots synthesized by the reflux method. Luminescence. 2015;30(4):465–471. doi: 10.1002/bio.2761. [DOI] [PubMed] [Google Scholar]
- 117.Huang D., Niu C., Zeng G., Wang X., Lv X. A highly sensitive protocol for the determination of Hg2+ in environmental water using time-gated mode. Talanta. 2015;132:606–612. doi: 10.1016/j.talanta.2014.10.015. [DOI] [PubMed] [Google Scholar]
- 118.Nanda Kumar D., Rajeshwari A., Alex S. A., Chandrasekaran N., Mukherjee A. Acetylcholinesterase inhibition-based colorimetric determination of Hg2+ using unmodified silver nanoparticles. New Journal of Chemistry. 2015;39(2):1172–1178. doi: 10.1039/c4nj01722f. [DOI] [Google Scholar]
- 119.Li W., Chen B., Zhang H., et al. BSA-stabilized Pt nanozyme for peroxidase mimetics and its application on colorimetric detection of mercury(II) ions. Biosensors and Bioelectronics. 2015;66:251–258. doi: 10.1016/j.bios.2014.11.032. [DOI] [PubMed] [Google Scholar]
- 120.Zhang Y., Jiang H., Wang X. Cytidine-stabilized gold nanocluster as a fluorescence turn-on and turn-off probe for dual functional detection of Ag+ and Hg2+ Analytica Chimica Acta. 2015;870(1):1–7. doi: 10.1016/j.aca.2015.01.016. [DOI] [PubMed] [Google Scholar]
- 121.Jia J., Chen H. G., Feng J., Lei J. L., Luo H. Q., Li N. B. A regenerative ratiometric electrochemical biosensor for selective detecting Hg2+ based on Y-shaped/hairpin DNA transformation. Analytica Chimica Acta. 2016;908:95–101. doi: 10.1016/j.aca.2015.12.028. [DOI] [PubMed] [Google Scholar]
- 122.Li Z., Xia S., Wang J., Bian C., Tong J. Determination of trace mercury in water based on N-octylpyridinium ionic liquids preconcentration and stripping voltammetry. Journal of Hazardous Materials. 2016;301:206–213. doi: 10.1016/j.jhazmat.2015.08.061. [DOI] [PubMed] [Google Scholar]
- 123.Somé I. T., Sakira A. K., Mertens D., Ronkart S. N., Kauffmann J.-M. Determination of groundwater mercury (II) content using a disposable gold modified screen printed carbon electrode. Talanta. 2016;152:335–340. doi: 10.1016/j.talanta.2016.02.033. [DOI] [PubMed] [Google Scholar]
- 124.Zahid A., Lashin A., Rana U. A., et al. Development of surfactant based electrochemical sensor for the trace level detection of mercury. Electrochimica Acta. 2016;190:1007–1014. doi: 10.1016/j.electacta.2015.12.164. [DOI] [Google Scholar]
- 125.Armas M. A., María-Hormigos R., Cantalapiedra A., Gismera M. J., Sevilla M. T., Procopio J. R. Multiparametric optimization of a new high-sensitive and disposable mercury (II) electrochemical sensor. Analytica Chimica Acta. 2016;904:76–82. doi: 10.1016/j.aca.2015.11.016. [DOI] [PubMed] [Google Scholar]
- 126.Shirzadmehr A., Afkhami A., Madrakian T. A new nano-composite potentiometric sensor containing an Hg2+-ion imprinted polymer for the trace determination of mercury ions in different matrices. Journal of Molecular Liquids. 2015;204:227–235. doi: 10.1016/j.molliq.2015.01.014. [DOI] [Google Scholar]
- 127.Xiong E., Wu L., Zhou J., Yu P., Zhang X., Chen J. A ratiometric electrochemical biosensor for sensitive detection of Hg2+ based on thymine-Hg2+-thymine structure. Analytica Chimica Acta. 2015;853(1):242–248. doi: 10.1016/j.aca.2014.10.015. [DOI] [PubMed] [Google Scholar]
- 128.Jing J., Yu L., Feng G. Z., Lei L. J., Qun L. H., Bing L. N. A regenerative electrochemical biosensor for mercury(II) by using the insertion approach and dual-hairpin-based amplification. Journal of Hazardous Materials. 2015;295:63–69. doi: 10.1016/j.jhazmat.2015.04.017. [DOI] [PubMed] [Google Scholar]
- 129.Chen D. M., Gao Z. F., Jia J., Li N. B., Luo H. Q. A sensitive and selective electrochemical biosensor for detection of mercury(II) ions based on nicking endonuclease-assisted signal amplification. Sensors and Actuators B: Chemical. 2015;210:290–296. doi: 10.1016/j.snb.2014.12.114. [DOI] [Google Scholar]
- 130.Ganjali M. R., Faridbod F., Davarkhah N., Shahtaheri S. J., Norouzi P. All solid state graphene based potentiometric sensors for monitoring of mercury ions in waste water samples. International Journal of Environmental Research. 2015;9(1):333–340. [Google Scholar]
- 131.Mao A., Li H., Cai Z., Hu X. Determination of mercury using a glassy carbon electrode modified with nano TiO2 and multi-walled carbon nanotubes composites dispersed in a novel cationic surfactant. Journal of Electroanalytical Chemistry. 2015;751:23–29. doi: 10.1016/j.jelechem.2015.04.034. [DOI] [Google Scholar]
- 132.Liao Y., Li Q., Wang N., Shao S. Development of a new electrochemical sensor for determination of Hg(II) based on Bis(indolyl)methane/Mesoporous carbon nanofiber/Nafion/glassy carbon electrode. Sensors and Actuators B: Chemical. 2015;215:592–597. doi: 10.1016/j.snb.2015.04.006. [DOI] [Google Scholar]
- 133.Mashhadizadeh M. H., Ramezani S., Rofouei M. K. Development of a novel MWCNTs-triazene-modified carbon paste electrode for potentiometric assessment of Hg(II) in the aquatic environments. Materials Science and Engineering C: Materials for Biological Applications. 2015;47:273–280. doi: 10.1016/j.msec.2014.11.011. [DOI] [PubMed] [Google Scholar]
- 134.Farahi A., Lahrich S., Achak M., Bakasse M., El Mhammedi M. A. Effect of KNO3 to remove silver interferences in the determination of mercury(II): application in milk and breast milk samples. Sensing and Bio-Sensing Research. 2015;4:90–95. doi: 10.1016/j.sbsr.2015.04.004. [DOI] [Google Scholar]
- 135.Farahi A., Achak M., ElGaini L., ElYamani S., Bakasse M., ElMhammedi M. A. Effect of pulse width of square potential to remove silver interference in the determination of mercury(II) Electroanalysis. 2015;27(8):1979–1988. doi: 10.1002/elan.201500029. [DOI] [Google Scholar]
- 136.Yang Y., Kang M., Fang S., et al. Electrochemical biosensor based on three-dimensional reduced graphene oxide and polyaniline nanocomposite for selective detection of mercury ions. Sensors and Actuators B: Chemical. 2015;214:63–69. doi: 10.1016/j.snb.2015.02.127. [DOI] [Google Scholar]
- 137.Lin Y., Peng Y., Di J. Electrochemical detection of Hg(II) ions based on nanoporous gold nanoparticles modified indium tin oxide electrode. Sensors and Actuators B: Chemical. 2015;220:1086–1090. doi: 10.1016/j.snb.2015.06.064.18641 [DOI] [Google Scholar]
- 138.Ting S. L., Ee S. J., Ananthanarayanan A., Leong K. C., Chen P. Graphene quantum dots functionalized gold nanoparticles for sensitive electrochemical detection of heavy metal ions. Electrochimica Acta. 2015;172:7–11. doi: 10.1016/j.electacta.2015.01.026. [DOI] [Google Scholar]
- 139.Fernández E., Vidal L., Martín-Yerga D., Blanco M. D. C., Canals A., Costa-García A. Screen-printed electrode based electrochemical detector coupled with ionic liquid dispersive liquid–liquid microextraction and microvolume back-extraction for determination of mercury in water samples. Talanta. 2015;135:34–40. doi: 10.1016/j.talanta.2014.11.069. [DOI] [PubMed] [Google Scholar]
- 140.Dong Z.-M., Qing X.-M., Zhao G.-C. Highly sensitive EQCM sensor for mercury determination by coupled stripping voltammetry. International Journal of Electrochemical Science. 2015;10(3):2602–2612. [Google Scholar]
- 141.Sabri Y. M., Kabir K. M. M., Boom E., Rosenberg S., Ippolito S. J., Bhargava S. K. Mercury detection in real industrial flue gas using a nanostructured quartz crystal microbalance. Industrial & Engineering Chemistry Research. 2016;55(28):7661–7668. doi: 10.1021/acs.iecr.6b01628. [DOI] [Google Scholar]
- 142.Kabir K. M., Sabri Y. M., Kandjani A. E., et al. Mercury sorption and desorption on gold: a comparative analysis of surface acoustic wave and quartz crystal microbalance-based sensors. Langmuir. 2015 doi: 10.1021/acs.langmuir.5b01858. [DOI] [PubMed] [Google Scholar]
- 143.Larki P., Sabri Y. M., Kabir K. M. M., Nafady A., Kandjani A. E., Bhargava S. K. Silver/gold core/shell nanowire monolayer on a QCM microsensor for enhanced mercury detection. RSC Advances. 2015;5(112):92303–92311. doi: 10.1039/c5ra19132g. [DOI] [Google Scholar]
- 144.Guo Z., Chen B., Wang Z., Jiang X. An electrochemiluminescence biosensor for mercury ion detection based on gamma-polyglutamic acid-graphene-luminol composite and oligonucleotides. Sensors and Actuators B: Chemical. 2015;209:579–585. doi: 10.1016/j.snb.2014.12.028. [DOI] [Google Scholar]
- 145.Cai F., Zhu Q., Zhao K., Deng A., Li J. Multiple signal amplified electrochemiluminescent immunoassay for Hg2+ using graphene-coupled quantum dots and gold nanoparticles-labeled horseradish peroxidase. Environmental Science & Technology. 2015;49(8):5013–5020. doi: 10.1021/acs.est.5b00690. [DOI] [PubMed] [Google Scholar]