Skip to main content
. 2016 Jul 8;13(4):353–360. doi: 10.1177/1479972316647177

Figure 5.

Figure 5.

Intermittent disconnection test in the volume-assisted setting with VT 500 mL. VT: delivered tidal volume; test periods (of 10 respiratory cycles): room air: mouthpiece disconnection; lung: mouthpiece connection to the test lung; obstr.: mouthpiece obstruction; *auto-triggering at mouthpiece disconnection.