Table 2. Sensitivity, specificity, and positive and negative predictive values of penicillinase tests using blaZ PCR as the standard.

Penicillinase test Sensitivity (%) Specificity (%) Positive predictive value (%) Negative predictive value (%)
Nitrocefin-based test 50 99.5 50 99.5
P2 zone diameter interpretation* 100 96.5 22.2 100
P2 zone edge test 100 99 50 100
P10 zone diameter interpretation 50 100 100 99.5
P10 zone edge test 100 95.5 18.2 100
Cloverleaf test 50 100 100 99.5

*P2 zone diameters were interpreted according to the EUCAST criteria (isolates with a P1 diameter of ≤25 mm were considered resistant) [11]; P10 zone diameters were interpreted according to the CLSI criteria (isolates with a P10 diameter of ≤28 mm were considered resistant) [11].