Skip to main content
. 2017 Feb 15;58(5):710–719. doi: 10.1093/jrr/rrw131

Fig. 6.

Fig. 6.

The time delay test of the internal/external tracking system. (a) The red line represents the Vicon system's output signal tracking AP motion, and the blue line represents the G-arm system's output signal. (b) A magnified picture of the first peak in (a). The time difference between the two first peaks is 1/30 s.