Table 1.

Precision and Accuracy Results for TFVdp, FTCtp, and 3TCtp in DBS Quality Control Samples

Analyte Nominal conc.
(fmol/sample)
Run 1
Accuracy
(%Bias)
Run 1
Precision
(%CV)
Run 2
Accuracy
(%Bias)
Run 2
Precision
(%CV)
Run 3
Accuracy
(%Bias)
Run 3
Precision
(%CV)
Inter-assay
Accuracy
(%Bias)
Inter-assay
Precision
(%CV)
TFVdp 100 −5.2 5.2 2.1 4.1 3.1 8.6 0.0 7.0
300 −4.0 5.5 0.3 2.6 −0.3 4.4 −1.3 4.5
2000 −1.5 2.5 −0.5 3.2 0.4 2.4 −0.5 2.7
20000 −1.6 2.3 −1.2 1.5 0.3 1.6 −0.8 1.9
FTCtp 100 −3.0 14.0 −0.2 4.4 2.9 6.7 −0.1 9.0
300 0.2 6.3 −1.8 3.7 −0.9 5.0 −0.9 4.9
2000 1.8 1.6 −1.0 1.5 −0.3 3.6 0.1 2.6
20000 −0.8 2.1 −3.4 1.0 0.2 2.4 −1.3 2.4
3TCtp 100 0.7 9.4 −9.6 8.0 1.8 8.4 −2.4 9.8
300 −2.1 5.8 −2.8 4.5 −3.9 5.2 −3.0 5.0
2000 0.1 1.5 −1.5 1.6 0.8 2.6 −0.2 2.1
20000 −0.5 1.2 −3.1 0.9 0.9 2.6 −0.9 2.4