TABLE 1.

Positive and negative agreement of seven treponemal assays with a consensus of the test panel

Assay and resulta Test panel consensus (no. of samples)b
% agreement (95% CI)
κ value
Positive Negative Positive Negative
BioPlex MBIA
    Positive 870 16 95.8 (94.3–97.0) 98.1 (97.1–98.8) 0.95
    Negative 31 1,066
    Indeterminate 7 5
Centaur CIA
    Positive 905 43 99.7 (99.0–99.9) 96.0 (94.7–97.1) 0.95
    Negative 2 1,044
    Indeterminate 1 0
FTA-ABS
    Positive 782 8 86.1 (83.7–88.2) 99.0 (98.2–99.5) 0.91
    Negative 82 1,076
    Indeterminate 44 3
Inno-LIA
    Positive 841 16 92.6 (90.7–94.2) 98.5 (97.6–99.1) 0.92
    Negative 67 1,071
Liaison CIA
    Positive 879 39 96.8 (95.4–97.8) 96.0 (94.6–97.0) 0.94
    Negative 20 1,043
    Indeterminate 9 5
TP-PA
    Positive 756 0 83.3 (80.7–85.6) 100 (99.6–100) 0.86
    Negative 137 1,087
    Indeterminate 15 0
Trep-Sure EIA
    Positive 902 149 99.3 (98.5–99.7) 86.3 (84.1–88.2) 0.85
    Negative 5 938
    Indeterminate 1 0
a

MBIA, microbead immunoassay; EIA, enzyme immunoassay; LIA, line immunoassay; TP-PA, Treponema pallidum particle agglutination assay; FTA-ABS, fluorescent treponemal antibody absorbed test; CIA, chemiluminescence immunoassay.

b

Consensus was defined as ≥4/7 tests positive or ≥4/7 tests negative (n = 1,995).