TABLE 6.
Selected functions more reflective of required microbiology mitigation controls in the traditional microbiology culture interpretation setting with percent compliance among WCLN risk assessment survey respondents, 2016
| Required mitigation control | No. of qualified respondents | % compliance |
|---|---|---|
| Perform aerosol-generating work in a biosafety cabinet with BSL-3 practices (high-risk bacterial pathogen) | 61 | 75.4 |
| Perform all work with AFB in a biosafety cabinet with BSL-3 practices | 17 | 76.5 |
| Perform all potential work with systemic fungi in a biosafety cabinet with BSL-3 practices | 22 | 54.5 |
| Perform all culture work with suspect or highly pathogenic viruses in a biosafety cabinet with BSL-3 practices | 3 | 33.3 |
| Post a sign warning others when working with systemic fungus cultures | 22 | 50.0 |
| Post a sign warning others when working with AFB cultures | 17 | 70.6 |
| Decontaminate waste (including liquid) from highly pathogenic microbes prior to disposal | 52 | 69.2 |