Table 1. A subset of assay parameters evaluated during validation.

Parameters Result
  AAV5 TAb assay AAV5 TI assay
Sensitivity (neat concentrations in plasma) 4.9 ng/ml of a polyclonal anti-AAV5 (or 68.9 ng/ml ADK5a equivalents) 65.9 ng/ml of a monoclonal anti-AAV5 (ADK5b)
Specificity No cross-reactivity of 4 μg/ml anti-AAV8 No cross-reactivity of 48 μg/ml anti-AAV8
Cut point 1.15 (SCP); 1.30 (TCP); 39.7% (CCP) 44.9% Transduction (ACP)
Interassay precision (%CV) HQC 14.3% TQC 3.78%
  LQC 13.1% LQC 17.5%
  NQC 9.9% NQC 35.6%
Intra-assay precision (%CV) HQC 4.6% TQC 3.29%
  LQC 7.2% LQC 11.6%
  NQC 6.3% NQC 13.2%
Selectivity 80% (8/10) pass at 0 ng/ml anti-AAV5 100% (10/10) pass at 0 ng/ml anti-AAV5
  100% (10/10) pass at 9 ng/ml anti-AAV5 100% (10/10) pass at 180 ng/ml anti-AAV5
Interference, hemolysis 100% (6/6) pass at 0 ng/ml anti-AAV5 66.7% (6/9) pass at 0 ng/ml anti-AAV5
  100% (6/6) pass at 9 ng/ml anti-AAV5 100% (9/9) pass at 180 ng/ml anti-AAV5
Interference, lipemia 66.7% (4/6) pass at 0 ng/ml anti-AAV5 33.3% (3/9) pass at 0 ng/ml anti-AAV5
  100% (6/6) pass at 9 ng/ml anti-AAV5 100% (9/9) pass at 180 ng/ml anti-AAV5

Abbreviations: %CV, percentage coefficient of variation; AAV, adeno-associated virus; ACP, assay cut point; CCP, confirmation or specificity cut point; HQC, high-quality control; LQC, low-quality control; NQC, negative quality control; SCP, screening cut point; TAb, total anti-capsid antibody; TCP, titer cut point; TI, transduction inhibition; TQC, titer quality control.