Table 3.
Column Specifications | Chromatographic Separation Conditions | ||||
---|---|---|---|---|---|
Name | Manufacturer | Stationary Phase | Dimensions | Flow (mL·min−1) | Elution Gradient (A: Acetonitrile; B: Water with Ammonium Acetate) |
XBridgeTM Amide | Waters (Milford, MA, USA) | BEH amide | 150 × 4.6 mm2 i.d., 5 μm | 0.15 | 0–4 min, at 5% B; 4–34 min, from 5% to 70% B; 34–42 min, at 70% B; and 42–44 min, at 5%B |
TSK Gel Amide-80 | Tosoh Bioscience (Tokyo, Japan) | Amide | 250 × 2.0 mm2 i.d., 5 μm | 0.15 | 0–3 min, at 5% B; 3–27 min, from 5% to 70% B; 27–30 min, at 70% B; and 30–32 min, at 5%B |
ZIC-HILIC | SeQuant (Umeå, Sweden) | Zwitterionic | 250 × 2.1 mm2 i.d., 5 μm | 0.15 | 0–3 min, at 5% B; 3–27 min, from 5% to 70% B; 27–30 min, at 70% B; and 30–32 min, at 5%B |
AcclaimTM Mixed-Mode HILIC-1 | Thermo Scientific (Sunnyvale, CA, USA) | Mixed-mode diol | 150 × 2.1 mm2 i.d., 5 μm | 0.15 | 0–2 min, at 5% B; 2–16 min, from 5% to 70% B; 16–20 min, at 70% B; and 20–22 min, at 5%B |