Skip to main content
. 2017 Dec 15;5(12):e198. doi: 10.2196/mhealth.8458

Table 4.

Comparison of Qardioarm and Omron devices in mean estimates and correlations of individual means between devices.


Variable
Omron, mean (SDa)
QardioArm, mean (SD)

t-test P value

r

P value
Intrasession 1





Systolic pressure (mm Hg) 127.34 (16.02) 126.47 (16.06) .70 .983 <.001

Diastolic pressure (mm Hg) 73.66 (9.82) 74.78 (10.29) .43 .984 <.001

Heart rate (bpm) 71.31 (10.29) 71.36 (10.42) .97 .987 <.001
Intrasession 2





Systolic pressure (mm Hg) 125.65 (16.02) 124.96 (15.84) .76 .984 <.001

Diastolic pressure (mm Hg) 72.23 (9.08) 73.90 (9.61) .21 .966 <.001

Heart rate (bpm) 70.12 (10.65) 70.05 (10.99) .96 .987 <.001
Intersession





Systolic pressure (mm Hg) 126.50 (15.47) 125.71 (15.30) .72 .991 <.001

Diastolic pressure (mm Hg) 72.95 (8.90) 74.34 (9.43) .29 .982 <.001

Heart rate (bpm) 70.72 (9.68) 70.70 (9.76) .99 .993 <.001

aSD: standard deviation.