Skip to main content
. 2017 Dec 27;11:11–20. doi: 10.2147/MDER.S142126

Table 2.

Subjects’ details (A) and observer measurement in each recruitment range (B) for the QardioArm® BP device according to the ESH-IP 2010

(A) Subject details: QardioArm®
Sex Male:female 18:15
Age (years) Range (low:high) 30:68
Mean (SD) 48.1 (10.5)
Arm circumference (cm) Range (low:high) 24:37
Mean (SD) 31.6(3.0)
Mercury Cuff Small 0
Standard 20
Large 13
Recruitment SBP DBP
BP (mmHg) Range (low:high) 101:180 49:129
Mean (SD) 143.2 (27.4) 91.2 (20.3)
(B) Observer Measurement in each Recruitment Range: Omron M6 IT Comfort®

SBP (mmHg)
Overall Range (low:high) 98:175
Low (<130) 36
Medium (130–160) 32
High (>160) 31
Maximum Difference 5
DBP (mmHg)
Overall range (low:high) 49:124
Low (<80) 33
Medium (80–100) 35
High (>100) 31
Maximum difference 4

Abbreviations: BP, blood pressure; DBP, diastolic blood pressure; ESH-IP, European Society of Hypertension International Protocol; SBP, systolic blood pressure; SD, standard deviation.