Table 3.

Parameters of variability and extraction recovery for 15d-PGJ2

Intraday Interday
Nominal concentration[pg mL−1] Recovery [%] Measured concentration [pg mL−1] Accuracy [%] Precision [% CV] Measured concentration [pg mL−1] Accuracy [%] Precision [% CV]
5 91.6 ± 4.6 5.3 ± 0.6 106.0 11.3 4.9 ± 0.3 98 6.1
150 88.6 ± 13.9 141.2 ± 11.5 93.9 8.1 136.7 ± 15.8 91.1 11.6
400 96.1 ± 6.1 461.9 ± 54.3 115.5 11.8 389.1 ± 57.2 97.3 14.7

Displayed is the intra- and interday accuracy/precision and the extraction efficiency based on the recovery. Accuracy is defined as the mean of the quantified concentration in percent of 3 spiked concentrations (nominal) in human artificial plasma samples. Precision is described as the CV of the mean concentration determined for 3 different concentrations (intraday, n = 4; interday, n = 4)